Fusion Ablation System recallZ-0831-2019 · AtriCure, Inc. · initiated Jan 22, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from AtriCure, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential for patients undergoing ablation to experience an elevated risk for TE; during stand-alone off-pump procedures where cardio pulmonary bypass (CPB) was not utilized.

FDA root cause category: Unknown/Undetermined by firm

What the company is doing

On January 22, 2019, AtriCure distributed URGENT Advisory Notices to customers via certified mail. AtriCure is providing modifications to the existing Cobra Fusion Instruction for Use (IFU), to address contributing TE Factors associated with stand-alone off-pump procedures. These are in addition to the warnings and precautions that are present within the current IFU. Modifications to Warnings: When utilizing the Cobra Fusion device in a stand-alone off-pump (without CPB) procedure, the following should be considered: - Physicians should consider a comprehensive anti-coagulation protocol including pre-operative, intra-operative and post-operative anti-coagulation management to prevent potential thromboemboli. Modifications to Precautions: - Post-operative anti-coagulation therapy for protection against thromboemboli is inclusive of the bridging period between the end of the procedure and until effective therapeutic levels of Oral Anticoagulation (OAC) are achieved. Customers Should Take the following Action: - Read and follow the revised Instructions for Use IFU enclosed with the notification letter when using the Cobra Fusion. - Report any post ablation thromboembolic event with the Cobra Fusion, to AtriCure by phone at 1-866-349-2342 (select option 6) or e-mail to [email protected]. - Return the completed Acknowledgement Form. Contact Information: If you have any questions, please contact Rob Cantu, Vice President of Quality at (1-513-644-4245) from 9-6pm ET on Mondays - Fridays. You may also contact customer service at (1-866-349-2342) any time of day, your message will be forwarded to Quality Assurance for review promptly. **Press Release will be issue depended on CDRH comment/review**

Recall details

Recall number
Z-0831-2019
Status
Terminated
Initiated by the company
Jan 22, 2019
Recalling firm
AtriCure, Inc.
Quantity
175 units
Product code
OCL
FDA submissions
K113475
Distribution
AL, AR, AZ, CA, CT, CO, FL, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA & WI Internationally - Germany, Italy, Netherlands, France, Austria, United Kingdom, Poland, Czech Republic, United Arab Emirates, Sweden, Switzerland, Kuwait, Belgium and Latvia

Product description

COBRA FUSION 150 Ablation System, Catalog # (REF) 700-001S

Code information (models, lots, serial numbers)

UDI - 00818354012828 Lot Numbers - All lots within expiry

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 81932).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page