Skip to content

DELFIA hCG Kit, In Vitro Diagnostic for the determination… recallZ-1237-2019 · PerkinElmer Life and Analytical Sciences, Wallac, OY · initiated Jan 10, 2019

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from PerkinElmer Life and Analytical Sciences, Wallac, OY or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

hCG analyte concentrations may be decreased up to 15- 20% at the lower measuring range of hCG concentrations.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Perkin Elmer initiated notification on January 10, 2018. The Recall Letter delivered to the end customer by email facilitated through the PerkinElmer Technical Support Group. PerkinElmer requires customers to: Inspect DELFIA / AutoDELFIA hCG kit inventory for the affected lots. Review your QC follow-up of the AutoDELFIA hCG assay for potential decrease in the control concentrations. If the control values are within your established acceptance range it is safe to report the results To avoid unnecessary repeat testing of samples, if the control values are out of the acceptance range discontinue use of the DELFIA / AutoDELFIA hCG kit lot and dispose the remaining kits of the lot. Complete the Response Form with the quantity of affected DELFIA / AutoDELFIA hCG kits you have disposed from your inventory

Recall details

Recall number
Z-1237-2019
Status
Open, Classified
Initiated by the company
Jan 10, 2019
Recalling firm
PerkinElmer Life and Analytical Sciences, Wallac, OY
Quantity
31 kits (10 in US, 21 in ROW)
Product code
JHJ
FDA submissions
K862044
Distribution
Worldwide distribution - US Nationwide distribution. and countries of Austria, Bolivia, UK, Switzerland,Thailand, Finland.

Product description

DELFIA hCG Kit, In Vitro Diagnostic for the determination of human gonadotropin (hCG) Product Code: A007-101 - Product Usage: Quantitative determination of human gonadotrophin (hCG) in serum.

Code information (models, lots, serial numbers)

Kit Lot Numbers: 664891, 666471, 666604 Pack Lot Numbers: 1066489101, 1066647101,1066660401 UDI:  (01)06438147009176(17)190430(10)664891 (01)06438147009176(17)190430(10)666471 (01)06438147009176(17)190430(10)666604

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 82064).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page