TriMed Hex Cortical Screw 4.0, 40 mm, REF HEX4.0-40… recallZ-1214-2019 · TriMed Inc. · initiated Aug 24, 2018
Class II Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from TriMed Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The bone screw is shorter than the intended design specification.
FDA root cause category: Component design/selection
What the company is doing
The recalling firm, TriMed, issued a "URGENT MEDICAL DEVICE RECALL" letter via email on 8/24/2018 to the distributors and third-party distributor. A follow-up letter was issued 3/28/2019 requesting downstream customers be notified to the user level. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: " If the customer further distributed Hex Cortical Screws, 4.0mm that is the subject of this recall, to forward the letter and the return response card to those entities that utilized the impacted product. " If the customer utilized the Hex Cortical Screws, 4.0mm that is the subject of this recall, then the customer is to: o Review customer inventory (including any kits that you have prepared for surgical operation) to see if they have any of the impacted product. o If a customer does have impacted product within its inventory, remove the product from use. o Review customer records to see if the impacted product was implanted. If the customer has records that impacted product was implanted, the customer should apply medical judgment to determine whether additional patient monitoring activities should be performed to evaluate implant placement, implant structural integrity, and fracture union. o Return to TriMed a return response card by 11 April 2019. Please email the response to: [email protected]. Please contact TriMed via telephone or email as per below if you have any questions: Telephone*: 800-633-7221 or Email: [email protected] * Hours of operation: 8:00AM to 5:00PM PST (Monday through Friday, excluding holidays).
Recall details
- Recall number
- Z-1214-2019
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Aug 24, 2018
- Recalling firm
- TriMed Inc.
- Quantity
- 13 screws
- Product code
- HRS
- FDA submissions
- K060041
- Distribution
- US Distribution to states of FL and MO.
Product description
TriMed Hex Cortical Screw 4.0, 40 mm, REF HEX4.0-40, non-Sterile, RX, a component of the TriMed Ankle Fixation System. The HEX4.0 40 bone screw is a component of the TriMed Ankle Fixation System. The HEX4.0-40 Screw is a single piece machined cortical bone screw. The HEX4.0-40 screw is employed in combination with a TriMed Ankle Bone Plate such as a Sidewinder Plate, Ankle Hook Plate, Cluster, Semi-Tubular Plate, or the Locking Plate. Use of these implants is not a substitute for normal tissue healing. The TriMed Ankle Bone Plates and Screws are designed to provide additional constraint of movement of a fractured bone and are intended only as an aid to fix the fracture in place during the healing process.
Code information (models, lots, serial numbers)
Lot #170281, UDI (01)00842188107108(11)170406(10)170281
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Hex Cortical Screw 4.0mm HEX4.0-40Trimed, INC. · Screw, fixation, bone1 recall0 reports
Other devices under the same FDA submission
4 devices cleared under K060041 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Locking Plates PS8Trimed, INC. · Plate, fixation, bone2 reports
- Locking Plates PCA8Trimed, INC. · Plate, fixation, bone1 report
- SideWinder Fracture Plates SWOR-7MTrimed, INC. · Plate, fixation, bone1 report
- Locking Cortical Screw 3.2mm LCBS3.2-10Trimed, INC. · Screw, fixation, bone1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page