Integrated ECG cable with 3-lead leadwires labeled as the… recallZ-1060-2019 · GE Healthcare, LLC · initiated Feb 22, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Integrated ECG cable with 3-lead leadwires can short circuit during defibrillation and conduct 25% of the defibrillation energy away from the patient.

FDA root cause category: Under Investigation by firm

What the company is doing

Consignees were initially notified via telephone call on about 02/22/2019 and followed up with an "URGENT MEDICAL DEVICE CORECTION" letter. Instructions included to immediately stop use of the affected Integrated ECG cables and quarantine any on hand. GE Healthcare will replace all affected cables at no cost and a GE Healthcare representative will contact each customer to arrange for the collection of the affected cables and provide replacement of new cables. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171.

Recall details

Recall number
Z-1060-2019
Status
Terminated
Initiated by the company
Feb 22, 2019
Recalling firm
GE Healthcare, LLC
Quantity
288 devices (23 US, 265 OUS)
Product code
IKD
Distribution
Worldwide Distribution - US Nationwide to AK, KS, MA, MA, ME, NJ, NJ, NM, NY, TX. Distributed worldwide to Chile, Finland, France, Germany, Italy, Korea (Republic Of), Russia, Spain, Sweden.

Product description

Integrated ECG cable with 3-lead leadwires labeled as the following: 1. ECG TRUNK CABLE, 3-LD W/ INTEGRATED GRABBER LEAD WIRE, AHA, 3.6 M/12 FT; 2. ECG TRUNK CABLE, 3-LD W/ INTEGRATED GRABBER LEAD WIRE, IEC, 3.6 M/12 FT Product Usage: The reusable ECG trunk cable enables ECG monitoring with GE monitoring equipment requiring rectangular 11-pin connector cable. These accessories are indicated for use by qualified medical personnel only.

Code information (models, lots, serial numbers)

1. Catalog Number 2106309-001, GTIN 00840682137928;   2. Catalog Number 2106309-002, GTIN 00840682137829