VITROS recallZ-1288-2019 · Ortho-Clinical Diagnostics · initiated Feb 20, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ortho-Clinical Diagnostics or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Luminometer Malfunction May Cause Inability to Process MicroWell Assays on VITROS Systems
FDA root cause category: Software design
What the company is doing
Ortho Clinical Diagnostics (Ortho) issued an "URGENT PRODUCT CORRECTION NOTIFICATION" , a customer letter (Ref. CL2019-052) dated 21 February 2019 via FedEx overnight courier and/or US Postal Service Priority Mail (for PO Boxes only) to customers who were shipped affected VITROS Systems/or whose VITROS System has a luminometer parts replacement. The customers were notified of the issue and instructed to notify Ortho upon experiencing the issue described: -Avoid performing a shutdown/restart on your systems unless directed by an Ortho representative or prompted by condition code help text. -If you shutdown/restart and encounter any of the above Luminometer condition codes and are unable to process MicroWell assays, immediately contact the Ortho Care" Technical Solutions Center for service to your system. -Complete the enclosed Confirmation of Receipt form no later than March 1, 2019 by Fax: 1.888.557.3759 or 1.585.453.4110 or e-Mail Address: [email protected]. -Please forward this notification if the product was distributed outside of your facility. Foreign affiliates were informed by email on 21 February 2019 of the issue and instructed to notify affected consignees of the issue and appropriate actions to take. Questions contact Ortho Care" Technical Solutions Center at 1-800-421-3311.
Recall details
- Recall number
- Z-1288-2019
- Status
- Terminated
- Initiated by the company
- Feb 20, 2019
- Recalling firm
- Ortho-Clinical Diagnostics
- Quantity
- 4 units
- Product code
- JJE
- FDA submissions
- K081543
- Distribution
- Worldwide Distribution: US (Nationwide) and countries of: Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden,The Netherlands, and United Kingdom.
Product description
VITROS 5600 Integrated System Refurbished-Software V3.3.2 & below Product Code: 6802915 For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents
Code information (models, lots, serial numbers)
Serial Numbers: Affects systems that had the Luminometer replaced during a service-repair UDI: 10758750007110
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- VITROS 6802915Ortho-Clinical Diagnostics, Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE8 recalls3 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 82311).
- Z-1287-2019 · TerminatedVITROS 5600 Integrated System -Software V3.3.2 & below Product Code: 6802413 For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety …
- Z-1289-2019 · TerminatedVITROS 3600 Immunoassay System-Software V3.3.2 & below Product Code: 6802783 For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of …
- Z-1290-2019 · TerminatedVITROS 3600 Immunoassay System Refurbished-Software V3.3.2 & below Product Code: 6802914 For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a…
- Z-1292-2019 · TerminatedVITROS ECi/ECiQ Immunodiagnostic System Product Code: 1922814 For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of cl…
- Z-1293-2019 · TerminatedVITROS ECi Immunodiagnostic System Refurbished Product Code: 6801059 For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page