UniCel DxH 600 Coulter Cellular Analysis System, REF number… recallZ-1383-2019 · Beckman Coulter Inc. · initiated Jul 30, 2018
Class I Terminated All recalls posted in May 2019
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Beckman Coulter Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Complaints received of sporadic erroneously elevated platelet results without flags or system messages. Thrombocytopenia may go unrecognized.
FDA root cause category: Software Design Change
What the company is doing
Urgent Medical Device Recall letters, dated July 30, 2018, were sent via email and postal mail on August 1, 2018 (for those customers not on their email notification system). On 05/22/19, second Urgent Medical Device (UMDC) Recall letters, and Important Safety Information notices, dated May 20, 2019 (replacing BEC UMDC letter dated July 30, 2018), were sent via mail and/or phone and/or e-mail informing customers perform the following actions: - Run samples on an instrument not subject to this recall to confirm the platelet results. - If an alternative instrument is not available, use the following quality control measures to aid in identification of discrepant platelet results: a) Perform manual scanning/estimate of platelets on a peripheral smear and compare with instrument results. Note that this method will identify samples with marked to moderate thrombocytopenia but may not identify smaller discrepancies. b) Repeat testing of samples in a workflow configuration may facilitate the identification of discrepancies. If an erroneous result is detected, review results from adjacent samples, i.e., those tested on the instrument both before and after the erroneous result. c) Additional instrument and/or LIS features including reference ranges, XM (exponentially-weighted moving average) and delta checks may be informative. d) Follow your laboratory's standard operating procedure to confirm unexpected results. - Ongoing investigation indicates that the probable root cause is the sweep flow disruption that may occur following the "Clear RBC Apertures" procedure. This potential root cause is currently under further investigation. Customers should discontinue using this procedure. If you suspect that your instrument has a clogged aperture that will not clear, discontinue use of the analyzer, contact Customer Support Center and request service. - Communicate to the ordering physicians the need to avoid patient treatment based solely on any single test resu
Recall details
- Recall number
- Z-1383-2019
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Jul 30, 2018
- Posted by the FDA
- May 23, 2019
- Recalling firm
- Beckman Coulter Inc.
- Quantity
- 2,059 devices
- Product code
- GKZ
- FDA submissions
- K140911
- Distribution
- Worldwide Distribution - US Nationwide including Puerto Rico. There was government and military distribution. Foreign distribution was made to Andorra, Australia, Austria, Belgium, Canada, China, Croatia, Egypt, France, Germany, Greece, Hungary, India, Israel, Italy, Japan, Korea, Kuwait, Latvia, Lithuania, Mexico, Morocco, Myanmar, Netherlands Antilles, New Zealand, Portugal, Romania, Russia, Saudi Arabia, Serbia, Slovenia, Spain, Switzerland, Trinidad and Tobago, U.A.E, United Kingdom, and Vietnam.
Product description
UniCel DxH 600 Coulter Cellular Analysis System, REF number B23858. Product Usage: Note: The DxH 600 hematology analyzer is the same instrument as the DxH 800 (K140911), except that it is a benchtop unit with no instrument cart, computer stand, and no connectivity capabilities. DxH 800 intended use states: The UniCel DxH 800 Analyzer is a quantitative multi-parameter, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The UniCel DxH 800 Analyzer identifies and enumerates the parameters indicated below on the following sample types: Whole Blood (Venous and Capillary) o WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV, NE%, NE#, LY%, LY#, MO%, MO#, EO%, EO#, BA%, BA#, NRBC%, NRBC#, RET%, RET#, MRV, IRF Pre-Diluted Whole Blood (Venous and Capillary) o WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV Body Fluids (cerebrospinal, serous and synovial) o TNC and RBC
Code information (models, lots, serial numbers)
All current software versions are impacted
Other devices under the same FDA submission
1 device cleared under K140911 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- DxH 900 Hematology System C11478Beckman Coulter, Inc. · Counter, Differential Cell1 recall17 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 82329).
- Z-1382-2019 · TerminatedUniCel DxH 800 Coulter Cellular Analysis System, REF numbers 629029, B24465, B24802, B68304, B66445, and B63322. Product Usage: The UniCel DxH 800 Analyzer is a quantitative mu…
- Z-1384-2019 · TerminatedUniCel DxH 900 Coulter Cellular Analysis System, REF number C11478 Product Usage: Note: The DxH 900 hematology analyzer is CLIA categorized under document CR180294 and is curre…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page