Elekta Unity systems recallZ-1297-2019 · Elekta Limited · initiated Apr 3, 2019

Class II Terminated All recalls posted in May 2019

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Elekta Limited or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is no warning in the Elekta Unity manual for the administration of gadolinium-based or other contrast agents for imaging or image guidance purposes has not been validated.

FDA root cause category: Labeling design

What the company is doing

Elekta notified customers on about 04/03/2019 via email with "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" letter. Instructions included to ensure all staff working with Elekta Unity are aware of the issue and recall information and post the recall notice in a place accessible to all users, e.g. Instructions for use, until the recall is closed. Additionally customer were asked to complete and return the Acknowledgement Form.

Recall details

Recall number
Z-1297-2019
Recall class
Class II
Status
Terminated
Initiated by the company
Apr 3, 2019
Posted by the FDA
May 8, 2019
Recalling firm
Elekta Limited
Quantity
9 devices
Product code
IYE
FDA submissions
K182076
Distribution
Worldwide Distribution - US Nationwide in the states of Texas and Wisconsin. Distribution internationally to Denmark, England, Germany, Italy, Netherlands

Product description

Elekta Unity systems Product Usage: Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan. Elekta Unity is intended for use with compatible Treatment Planning and Oncology Information Systems.

Code information (models, lots, serial numbers)

Product Code: UNITY SYSTEM/13553-006/600003, UNITY SYSTEM/10817-011/600010, UNITY SYSTEM/10562-011/600016; UNITY SYSTEM/10719-T03001/600013, UNITY SYSTEM/30004070-001/600023, UNITY SYSTEM/10420-AVL-U/600008, UNITY SYSTEM/11014-UMCU-U/600007, UNITY SYSTEM/11611-40/600011, UNITY SYSTEM/11489-21/600009;   UDI: 1536549, 1536549

Other devices under the same FDA submission

1 device cleared under K182076 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.