LimiTorr INS9020 INS-9020 LIMITORR VOLUME LIMITING EVD 20 ML recallZ-1312-2019 · Integra LifeSciences Corp. · initiated Apr 8, 2019
Class I Terminated All recalls posted in May 2019
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Integra LifeSciences Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The firm has identified a complaint trend regarding breakage of the LimiTorr Transducer Mount stopcock and MoniTorr Panel Mount stopcock.
FDA root cause category: Device Design
What the company is doing
On April 8, 2019, the firm distributed Urgent Medical Device Correction letters to its affected customers. The communication advised customers of the possible breakage of the stopcock. Customers were advised to take the following precautions: When connecting a fluid-filled transducer (FFT) to the Panel Mount/Transducer Mount stopcocks: " Do not apply excessive force while attaching the FFT to the stopcock. " Take care during the priming process to not torque or bend the syringe and FFT. " Ensure luer lock is "finger tight;" do not over-tighten connection at the stopcock junction. When managing an external ventricular or lumbar drain (EVD) with a FFT attached: " Take care when manipulating the stopcock to zero the FFT and to obtain pressure readings, do not apply excessive force to the stopcock junction. " Use caution during patient transportation and during routine ICU care, to avoid direct contact of the FFT with other devices or equipment. " If the FFT must be flushed or replaced, do not apply excessive torque or force to the FFT/ stopcock junction. " If the FFT is connected to the ICU bedside monitor, ensure the interface cable is positioned so that it does not place excess force on the FFT (e.Q. do not allow cable to touch the floor. Customers were asked to determine if they had any affected product on hand, review and understand the precautions, and complete and return a completed acknowledgement form to the firm. Distributors were asked to forward the customer letter to any customers who received the affected product. Should you have any questions regarding these instructions, please contact Customer Service Monday to Friday 8:00 AM - 8:00 PM EST USA: 1-800-654-2873: email: [email protected] Outside USA: 781-565-1401 ***UPDATE*** On October 22, 2019, the firm sent out an updated notification informing customers that a new stopcock has been designed for the LimiTorr and MoniTorr, and is now available for replacement of th
Recall details
- Recall number
- Z-1312-2019
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Apr 8, 2019
- Posted by the FDA
- May 20, 2019
- Recalling firm
- Integra LifeSciences Corp.
- Quantity
- 42102 units
- Product code
- JXG
- FDA submissions
- K072929
- Distribution
- Worldwide Distribution - US Nationwide International distribution to Argentina, American Samoa, Australia, Canada, and Guam.
Product description
LimiTorr INS9020 INS-9020 LIMITORR VOLUME LIMITING EVD 20 ML LimiTorr INS9020SP1 INS-9020SP1 LIMITORR VOLUME EVD 20ML LimiTorr INS9030 INS-9030 LIMITORR VOLUME LIMITING EVD 30 ML LimiTorr INS9030SP1 INS-9030SP1 LIMITORR VOLUME LIMITING EVD 30 ML Product Usage: The LimiTorr External CSF Drainage and Monitoring Systems provides a closed system for the CSF from the ventricles of the brain or the lumbar subarachnoid space to an external drainage bag. The LimiTorr volume limiting external CSF drainage system was designed to include a volume limiting valve mechanism which reduces the chances for excessive CSF drainage. The burette in the volume limiting external CSF drainage system contains a volume limiting valve which stops drainage when the pre-determined volume (20ml-30ml) is reached. Indications: The LimiTorr system allows for drainage and monitoring of CSF from the lateral ventricles of the brain and the lumbar subarachnoid space in selected patients to reduce intracranial pressure (ICP), to monitor CSF, to provide temporary drainage of CSF in patients with infected CSF shunts, and to monitor ICP.
Code information (models, lots, serial numbers)
All lots
Other devices under the same FDA submission
2 devices cleared under K072929 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- LimiTorr™ INS9030Integra Lifesciences Corporation · Shunt, central nervous system and components2 reports
- LimiTorr™ INS9020Integra Lifesciences Corporation · Shunt, central nervous system and components1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 82659).
- Z-1310-2019 · TerminatedMoniTorr 10100 10-100 CSF Drainage System w/Patient Line One Way Valve MoniTorr 10102 10-102 CSF Drainage System w/Patient Line One Way Valve w MoniTorr 10110 10-110 CSF Drainage…
- Z-1311-2019 · TerminatedMoniTorr INS1100 INS-1100 CSF Drainage System used with Pole Mount System Product Usage: The MoniTorr ICP External CSF Drainage and Monitoring Systems provide a simple to use, …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page