TactiSys recallZ-1493-2019 · Abbott Laboratories Inc. (St Jude Medical) · initiated Apr 5, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Laboratories Inc. (St Jude Medical) or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

In reported cases, the device log on the TactiSys Quartz Equipment operating on Software Version 1.7.0 fills the allocated disk space, which prevents the storage of new log data. This may lead to intermittent contact force data to be displayed during the procedure.

FDA root cause category: Software design

What the company is doing

The recall notice, and clinician acknowledgment form were hand delivered by Abbott Sales Representatives to all impacted accounts starting April 5, 2019. Completed Clinician Acknowledgement Forms, signify they received the correction notification for the impacted TactiSys Quartz Equipment with software version 1.7.

Recall details

Recall number
Z-1493-2019
Status
Terminated
Initiated by the company
Apr 5, 2019
Recalling firm
Abbott Laboratories Inc. (St Jude Medical)
Quantity
317 units
Product code
OAE
FDA submissions
P130026
Distribution
US, UNITED KINGDOM, UAE, TURKEY, THAILAND, TAIWAN, SWITZERLAND, SWEDEN, SPAIN, SOUTH AFRICA, SLOVENIA, SLOVAKIA, SINGAPORE, SERBIA, SAUDIA ARABIA, RUSSIA, ROMANIA, PORTUGAL, POLAND, PALESTINE, OMAN, NORWAY, NEW ZEALAND, NETHERLANDS, MOROCCO, MALAYSIA, LEBANON, JORDAN, JAPAN, ITALY, ISRAEL, IRELAND, IRAN, GREECE, GERMANY, FRANCE, EL SALVADOR, EGYPT, DENMARK, Czech Republic, CHINA, CANADA, BULGARIA, BELGIUM, AUSTRIA, AUSTRALIA

Product description

TactiSys Quartz Equipment (with software version 1.7) Model: PN-004 400 Part Numbers: 600043767 (China), 100154367 (Global), 100154368 (SWAP)

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

UDI: 07640157990194 Serial Numbers: 33019, 33059, 33350, 33354, 33357, 33360, 33361, 33363, 33390, 33392, 33393, 33405, 33417, 33421, 33444, 33449, 33451, 33457, 33458, 33461, 33491, 33492, 33509, 33013, 33015, 33020, 33040, 33075, 33091, 33097, 33160, 33197, 33384, 33385, 33387, 33389, 33414, 33433, 33388, 33435, 33481, 33154, 33157, 33162, 33163, 33278, 33092, 33159, 33437, 33412, 33014, 33027, 33028, 33029, 33038, 33049, 33057, 33082, 33096, 33192, 33222, 33268, 33379, 33383, 33397, 33398, 33399, 33400, 33422, 33423, 33478, 33320, 33352, 33395, 33351, 33069, 33381, 33455, 33064, 33093, 33313, 33337, 33161, 33426, 33083, 33321, 33420, 33051, 33155, 33196, 33256, 33353, 33460, 33217, 33427, 33073, 33150, 33076, 33148, 33434, 33396, 33050, 33333, 33333, 33336, 33336, 33095, 33143, 33143, 33112, 33113, 33266, 33267, 33269, 33270, 33271, 33272, 33273, 33274, 33275, 33276, 33283, 33287, 33288, 33291, 33325, 33356, 33359, 33362, 33404, 33459, 33016, 33098, 33099, 33100, 33101, 33102, 33103, 33104, 33105, 33106, 33107, 33108, 33110, 33111, 33127, 33130, 33131, 33265, 33018, 33024, 33025, 33087, 33152, 33156, 33220, 33230, 33231, 33254, 33279, 33281, 33315, 33416, 33428, 33438, 33439, 33442, 33443, 33463, 33330, 33386, 33228, 33033, 33048, 33149, 33259, 33282, 33464, 33022, 33031, 33035, 33037, 33044, 33045, 33046, 33055, 33066, 33067, 33068, 33078, 33080, 33085, 33086, 33088, 33094, 33129, 33137, 33138, 33140, 33144, 33158, 33195, 33218, 33221, 33229, 33253, 33257, 33262, 33280, 33289, 33293, 33326, 33328, 33355, 33364, 33382, 33403, 33406, 33408, 33409, 33411, 33424, 33431, 33440, 33445, 33446, 33448, 33450, 33454, 33462, 33465, 33466, 33169, 33011, 33012, 33021, 33023, 33034, 33054, 33056, 33074, 33079, 33089, 33090, 33109, 33191, 33193, 33194, 33215, 33224, 33314, 33316, 33327, 33329, 33358, 33375, 33376, 33407, 33415, 33419, 33429, 33447, 33219, 33258, 33292, 33334, 33335, 33032, 33410, 33081, 33170, 33171, 33172, 33173, 33198, 33199, 33201, 33203, 33204, 33205, 33206, 33207, 33208, 33209, 33210, 33211, 33212, 33213, 33214, 33294, 33295, 33296, 33298, 33299, 33300, 33301, 33302, 33303, 33304, 33365, 33366, 33367, 33368, 33369, 33370, 33371, 33372, 33373, 33374, 33549, 33550, 33551, 33168, 33165, 33216, 33042, 33456, 33070, 33432, 33480, 33047, 33072, 33077, 33136, 33226, 33263, 33377, 33378,