Skip to content

Opteform Allograft Disc, 30mm 3mm, 2cc, Catalog Number… recallZ-1688-2019 · Exactech, Inc. · initiated Apr 26, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Exactech, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Opteform Disc-30mm, Opteform Disc-45mm, and Opteform Disc-90mm were potentially exposed to higher temperatures than those documented in the IFU due to a failure of the cold storage equipment.

FDA root cause category: Environmental control

What the company is doing

The firm initiated the recall on 04/26/2019 by electronic mail. The notice requested the consignee cease distribution of the product, notify their customers (the user), quarantine all units subject to recall in inventory, and if any units were implanted to follow-up on the patient's condition 30 days after surgery.

Recall details

Recall number
Z-1688-2019
Status
Terminated
Initiated by the company
Apr 26, 2019
Recalling firm
Exactech, Inc.
Quantity
2 units
Product code
MQV
FDA submissions
K043421
Distribution
OK, VA, FL

Product description

Opteform Allograft Disc, 30mm 3mm, 2cc, Catalog Number 600-03-30

Code information (models, lots, serial numbers)

UDI 10885862093202 Serial Numbers: T31634510, T31788559

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

2 devices cleared under K043421 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 82759).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page