Skip to content

REVACLEAR recallZ-1909-2019 · Baxter Healthcare Corporation · initiated May 31, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential presence of ruptured dialyzer fibers which may lead to a blood leak during treatment.

FDA root cause category: Process control

What the company is doing

The firm, Baxter, sent an " Urgent Product Recall" letter on 05/31/2019. The notice described the product, problem and actions to be taken. The customers were instructed to do the following: 1. Locate and remove all affected product lots from your facility. The product code and lot number can be found on the individual product or shipping carton. 2. Contact Baxter Healthcare Center for Service to arrange for return and credit. Baxter Healthcare Center for Service can be reached at 888-229-0001 between the hours of 7:00 am and 6:00 pm Central Time, Monday through Friday. Please have your Baxter 8-digit ship-to account number, product code, lot number, and quantity of product to be returned ready when calling. 3. If you purchased this product directly from Baxter, complete the enclosed Baxter customer reply form and return it to Baxter by faxing it to 224-270-5457 or scanning and e-mailing it to [email protected], even if you do not have any inventory. Returning the customer reply form promptly will confirmyour receipt of this notification and prevent you from receiving repeat notices. 4. If you purchased this product from a distributor, please note that the Baxter customer reply form is not applicable. If a reply form is provided by your distributor or wholesaler, please return it to the supplier according to their instructions. 5. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please conduct a consumer-level recall of the affected product that you distributed to customers. For general questions regarding this communication, contact Baxter Corporate Product Surveillance at 800-437-5176, between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday. .

Recall details

Recall number
Z-1909-2019
Status
Terminated
Initiated by the company
May 31, 2019
Recalling firm
Baxter Healthcare Corporation
Quantity
65403 units
Product code
KDI
FDA submissions
K130039
Distribution
US Nationwide Distribution and Internationally to: Canada and Bermuda.

Product description

REVACLEAR 400 Dialyzer, Product Code 114746L Revaclear 300 and Revaclear 400 dialyzers are indicated for treatment of chronic and acute renal failure by hemodialysis.

Code information (models, lots, serial numbers)

UDI 07332414124076 Lot Numbers: C419205101 C419205201 C419205301

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 83015).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page