Zyno Medical Administration Set -105"ADMIN SET 20DP, W/2 Y… recallZ-2058-2019 · Zyno Medical LLC · initiated May 29, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zyno Medical LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified

FDA root cause category: Process control

What the company is doing

On May 29, 2019 Zyno issued letter to the Distributor advising of the problem and health risk and action to take: remove all recalled products with codes, cease use and immediately and return them to Zyno Medical for replacement from unaffected lots. Call or email for instructions on the return of products: (508) 907-7772 (Monday to Friday, 9:30 a.m. to 5 p.m. EST) or [email protected] to acknowledge receiving this recall notice. Zyno issued a follow-up recall letter on 6/10/19 via email to the Distributor to issue to the user/customer level.

Recall details

Recall number
Z-2058-2019
Status
Terminated
Initiated by the company
May 29, 2019
Recalling firm
Zyno Medical LLC
Quantity
354 cases (50/case)=17700
Product code
FPA
FDA submissions
K132841
Distribution
MA

Product description

Zyno Medical Administration Set -105"ADMIN SET 20DP, W/2 Y SITE, CVCLMPS- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code:B2-70072

Code information (models, lots, serial numbers)

Lot Numbers:18037128,18065375, 19015969 UDl/UPC/GTIN: 00814371020198

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

13 devices cleared under K132841 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 83026).