SHERPA NX ACTIVE recallZ-2627-2019 · Medtronic Vascular · initiated Mar 15, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Vascular or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for extensive loss of primary segment material exposing underlying stainless-steel braid wires on a subset of 6F Sherpa catheters.
FDA root cause category: Component design/selection
What the company is doing
Medtronic initiated immediate communications to consignees on 15-Mar-2019. Consignees were asked to quarantine and/or return unused affected product, and confirmation of the immediate notification was requested per local requirements. In the US, Medtronic representatives notified consignees. Consignees were asked to return a Customer Confirmation Certificate. On 11-Apr-2019, Medtronic initiated the US follow up communications via hand-delivery and UPS 2-day delivery mailings to the March 2019 notification consignees and risk managers identified as part of the April 2019 scope expansion. Consignees are asked to quarantine and return unused affected product. The consignees are asked to return a confirmation form, acknowledging they have received the notification and have taken the requested action. Beginning 10-May-2019, Medtronic initiated additional communications via UPS 2-day delivery to 23 consignees in the US who were mailed the April 2019 notification with inaccurate product inventory reports. This communication included updated customer inventory reports. Communication was also initiated via UPS 2-day delivery to one consignee who did not receive the April 2019 notification, but did receive affected product. The consignees were asked to quarantine and return unused affected product. The consignees are asked to return a confirmation form, acknowledging they have received the notification and have taken the requested action. Beginning 10-Jun-2019, communication and retrieval activities will be initiated with all consignees who have received affected product (original and expanded scope) according to Medtronic records. Consignees will be asked to quarantine and return all unused affected product within their possession. Consignees will be asked to confirm receipt of this June 2019 FCA notification even if they have already confirmed notification of previous communications.
Recall details
- Recall number
- Z-2627-2019
- Status
- Terminated
- Initiated by the company
- Mar 15, 2019
- Posted by the FDA
- Oct 1, 2019
- Recalling firm
- Medtronic Vascular
- Quantity
- 167 units
- Product code
- DQY
- FDA submissions
- K042489, K132673
- Distribution
- Worldwide distribution to US, Austria, Belgium, Bulgaria, Czech Republic, France, Germany, Greece, Hong Kong, Ireland, Italy, Kazakhstan, Luxembourg, Moldova, Netherlands, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Taiwan, and United Kingdom
Product description
SHERPA NX ACTIVE GUIDING CATHETER, 6F ERAD LEFT, .070", REF SA6ERADL. for cardiovascular use
Code information (models, lots, serial numbers)
GTIN: 00613994824769 All Lot/Serial Numbers
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- SHERPA NX ACTIVE SA6ERADLMedtronic, Inc. · CATHETER, PERCUTANEOUS1 recall0 reports
Product lines
Other devices under the same FDA submission
194 devices cleared under K042489, K132673 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- SHERPA NX ACTIVE SA6EBU375Medtronic, Inc. · CATHETER, PERCUTANEOUS2 recalls0 reports
- SHERPA NX ACTIVE SA6IMAKMedtronic, Inc. · CATHETER, PERCUTANEOUS1 recall5 reports
- SHERPA NX ACTIVE SA6JR40SHMedtronic, Inc. · CATHETER, PERCUTANEOUS1 recall1 report
- SHERPA NX ACTIVE SA6EBU45Medtronic, Inc. · CATHETER, PERCUTANEOUS1 recall1 report
- SHERPA NX ACTIVE SA6RDND1KMedtronic, Inc. · CATHETER, PERCUTANEOUS1 recall1 report
- SHERPA NX ACTIVE SA6LCBSHDMedtronic, Inc. · CATHETER, PERCUTANEOUS1 recall0 reports
- SHERPA NX ACTIVE SA6LCBSHMedtronic, Inc. · CATHETER, PERCUTANEOUS1 recall0 reports
- SHERPA NX ACTIVE SA6LCBDMedtronic, Inc. · CATHETER, PERCUTANEOUS1 recall0 reports
- SHERPA NX ACTIVE SA6LCBMedtronic, Inc. · CATHETER, PERCUTANEOUS1 recall0 reports
- SHERPA NX ACTIVE SA6MAC30Medtronic, Inc. · CATHETER, PERCUTANEOUS1 recall0 reports
- SHERPA NX ACTIVE SA6MAC3030Medtronic, Inc. · CATHETER, PERCUTANEOUS1 recall0 reports
- SHERPA NX ACTIVE SA6MAC30SHMedtronic, Inc. · CATHETER, PERCUTANEOUS1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 82525).
- Z-2562-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, IMA 55CM .070", REF SA6IMAK. for cardiovascular use
- Z-2563-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, RDND1 55CM .070", REF SA6RDND1K. for cardiovascular use
- Z-2564-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, 3DRC 100CM .070", REF SA63DRC. for cardiovascular use
- Z-2565-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, AL1.0 110CM .070", REF SA6AL10A. for cardiovascular use
- Z-2566-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, AL1.0 90CM.070", REF SA6AL10D. for cardiovascular use
- Z-2567-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, AL3.0 110CM.070", REF SA6AL30A. for cardiovascular use
- Z-2568-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, AL.75 110CM.070", REF SA6AL75A. for cardiovascular use
- Z-2569-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, AR1.0 100CM.070", REF SA6AR10. for cardiovascular use
- Z-2570-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, AR1.0SH 100CM.070", REF SA6AR10SH. for cardiovascular use
- Z-2571-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, AR2.0, 100CM.070", REF SA6AR20. for cardiovascular use
- Z-2572-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, AR2.0 SH, 100CM.070", REF SA6AR20SH. for cardiovascular use
- Z-2573-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, CHAMP0.5, 100CM.070", REF SA6CHAMP05. for cardiovascular use
- Z-2574-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, CHAMP0.5 SH, 100CM.070", REF SA6CHAMP05SH. for cardiovascular use
- Z-2575-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, CHAMP2.0, 55CM.070", REF SA6CHAMP20K. for cardiovascular use
- Z-2576-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, EBU3.5, 110CM.070", REF SA6EBU35A. for cardiovascular use
- Z-2577-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, EBU3.5, 90CM.070", REF SA6EBU35D. for cardiovascular use
- Z-2578-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, EBU4.0, 110CM.070", REF SA6EBU40A. for cardiovascular use
- Z-2579-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, FL4.0, 100CM.070", REF SA6FL40. for cardiovascular use
- Z-2580-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, FL4.0 SH, 100CM, 070", REF SA6FL40SH. for cardiovascular use
- Z-2581-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, HSI, 100CM, 070", REF SA6HSI. for cardiovascular use
- Z-2582-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, HSI SH, 100CM, 070", REF SA6HSISH . for cardiovascular use
- Z-2583-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, IMA, 100CM, 070", REF SA6IMA. for cardiovascular use
- Z-2584-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, IMA, 90CM, 070", REF SA6IMAD . for cardiovascular use
- Z-2585-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, IMA SH, 100CM, 070", REF SA6IMASH. for cardiovascular use
- Z-2586-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, IMA SH, 60CM, 070", REF SA6IMASHJ. for cardiovascular use
- Z-2587-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, JCR4.0, 100CM, 070", REF SA6JCR40. for cardiovascular use
- Z-2588-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, JL3.5, 110CM, 070", REF SA6JL35A. for cardiovascular use
- Z-2589-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, JL4.0, 110CM, 070", REF SA6JL40A. for cardiovascular use
- Z-2590-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, JL4.0, 90CM, 070", REF SA6JL40D. for cardiovascular use
- Z-2591-2019 · TerminatedSHERPA NX ACTIVE GUIDING CATHETER, 6F, JL5.0, 110CM, 070", REF SA6JL50A. for cardiovascular use
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page