PTS Panels Lipid Panel test strip: IVD Lipid Panel Test… recallZ-2157-2019 · Polymer Technology Systems, Inc. · initiated Jun 7, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Polymer Technology Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The MEMo chips provided with lot P901 may be mislabeled, the MEMo Chip may display Q901 on the analyzer LCD screen, instead of P901
FDA root cause category: Under Investigation by firm
What the company is doing
PTS Diagnostic issued Medical Device Recall issued letter on June 7, 2019 via certified mail, letter provides reason for recall, health risk and action to take: Immediately check your inventories and, if you have further distributed this product, please identify the users and notify them at once of this product recall. 1. Immediately check your inventories and segregate any remaining materials from lot P901.2. Distributors should immediately forward this notice and response form to all customers/users who may have received this product. 3. Acknowledge receipt of this instruction by completing the attached form, even if you do not have affected product on hand.4. If you have remaining stock of lot P901, hold that stock in quarantine. Our Customer Care team will contact you to arrange for return of the materials.
Recall details
- Recall number
- Z-2157-2019
- Status
- Terminated
- Initiated by the company
- Jun 7, 2019
- Recalling firm
- Polymer Technology Systems, Inc.
- Quantity
- 1,887 vials US; 2,728 vials OUS
- Product code
- CHH
- FDA submissions
- K151545
- Distribution
- Worldwide Distribution - US Nationwide Bahamas Belgium Bermuda Brazil England Indonesia Latvia Mexico Nigeria Puerto Rico South Africa Sweden Taiwan
Product description
PTS Panels Lipid Panel test strip: IVD Lipid Panel Test Strips Ref Number: 1710 Product Usage: PTS Panels Lipid Panel test strips measure total cholesterol, HDL (high density lipoprotein) cholesterol, and triglycerides in whole blood with the CardioChek PA or the CardioChek Plus professional analyzers, and provide a quantitative result. A TC/HDL ratio and estimated values for LDL (low density lipoprotein) cholesterol are calculated by the CardioChek PA and CardioChek Plus analyzers. A MEMo Chip is provided with each package of test strips and must be properly inserted into the analyzer before any test can be run. The MEMo Chip contains the test name, calibration curve, lot number and test strip expiration date. After the test strip is inserted into the analyzer and blood applied to the test strip, test results are displayed in as little as 90 seconds.
Code information (models, lots, serial numbers)
Lot Number: P901 Exp. Date: 8/17/2020 (01) 00381931710016 (17) 200817 (10) P901
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- PTS Panels® Lipid Panel Test Strips 1710Polymer Technology Systems, Inc. · COLORIMETRIC METHOD, TRIGLYCERIDES1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 83097).
- Z-2158-2019 · TerminatedHenry Schein Panels Lipid Panel test strip: IVD Lipid Panel Test Strips Ref Number: 570-0414 Product Usage: PTS Panels Lipid Panel test strips measure total cholesterol, HDL (hi…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page