Spacelabs Healthcare Smart Disclosure System, Model 92810… recallZ-2035-2019 · Spacelabs Healthcare, Inc. · initiated Jan 28, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Spacelabs Healthcare, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Several reports were received that patient records were printed with the correct patient demographics but containing another patient's waveforms.
FDA root cause category: Software design
What the company is doing
The recalling firm began issuing customer letters on 1/28/2019 via priority service, return receipt requested, to all U.S. customers. The letter was translated, as necessary, and emailed on 2/4/2019 to the international subsidiaries and distributors. The firm identified 6 new customers (5 U.S. and 1 international) and the letter for the 5 new U.S. customers was issued 5/30/2019 via priority service, return receipt requested. The letter for the 1 new international customer was emailed on 5/28/2019.
Recall details
- Recall number
- Z-2035-2019
- Status
- Terminated
- Initiated by the company
- Jan 28, 2019
- Recalling firm
- Spacelabs Healthcare, Inc.
- Quantity
- 295 customer sites have the affected software versions
- Product code
- DQK
- FDA submissions
- K110779
- Distribution
- Distribution was nationwide. There was also government and military distribution. Foreign distribution was made to Argentina, Bahrain, Bolivia, Brazil, Canada, Cayman Islands, Chile, China, Costa Rica, Czech Republic, France, Hong Kong, India, Indonesia, Italy, Kuwait, Malaysia, Mexico, Netherlands, Pakistan, Palestine, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, Spain, Switzerland, Taiwan, Trinidad and Tobago, Tunisia, Turkey, UAE, United Kingdom, and Viet Nam.
Product description
Spacelabs Healthcare Smart Disclosure System, Model 92810, a component of the Intesys Clinical Suite (ICS) G2.
Code information (models, lots, serial numbers)
All affected Smart Disclosure installations with software versions 5.00, 5.01, 5.02, or 5.03, UDI 10841522123580.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Intesys® Clinical Suite Smart DisclosureSpacelabs Healthcare (Washington), Inc · COMPUTER, DIAGNOSTIC, PROGRAMMABLE1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page