Skip to content

VITROS recallZ-2202-2019 · Ortho-Clinical Diagnostics · initiated Jun 17, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ortho-Clinical Diagnostics or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potentially Biased Results using VITROS Chemistry Products GLU Slides

FDA root cause category: Process control

What the company is doing

Ortho issued On 17 June 2019, a customer letter (Ref. CL2019-172) via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to customers and distributors. The customers were notified of the issue and instructed: Discontinue using Lot 0048-0964-9287, discard any remaining inventory and switch to an alternate lot. " If you do NOT have an alternate lot, consider the following alternate options: o Use an alternate glucose method until your replacement order arrives. o Consult with your medical director to determine if the bias observed and rate of occurrence when using the affected lot is acceptable for continued use, and if so, continue to use until your replacement arrives. " Ortho will credit and/or replace your remaining inventory of Lot 0048-0964-9287. Please specify full and/or partial sales units remaining and indicate your preference of credit or replacement on your Confirmation of Receipt form. " Complete the enclosed Confirmation of Receipt form no later than June 21, 2019. " Post this notification by each system that processes VITROS GLU Slides or with your user documentation until you switch to an alternate lot. " Please forward this notification if the product was distributed outside of your facility. If you have further questions, please contact Ortho Care" Technical Solutions Center at 1-800-421-3311.

Recall details

Recall number
Z-2202-2019
Status
Terminated
Initiated by the company
Jun 17, 2019
Recalling firm
Ortho-Clinical Diagnostics
Quantity
1099 (5 pack x 60)=329700 slides
Product code
CGA
FDA submissions
K182072, K955286
Distribution
US Nationwide Distribution - AL, AR, CA, CO, CT, FL, KS, KY, MA, MN, TN

Product description

VITROS Chemistry Products GLU Slides-Product Code:1707801. IVD VITROS Chemistry Products GLU Slides quantitatively measure glucose (GLU) concentration in serum, plasma, urine, and cerebrospinal fluid using VITROS 250/350/5,1 FS/4600 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems

Code information (models, lots, serial numbers)

Lot Number:0048-0964-9287; Exp Date:2020-11-01 UDI: 10758750009572

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page