HS70A Diagnostic Ultrasound System Version 2.01.00… recallZ-2275-2019 · NeuroLogica Corporation · initiated Jul 25, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from NeuroLogica Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for probes overheating when decreasing the Doppler SV (Sample Volume) size value set by the user in the Doppler Only mode.
FDA root cause category: Software design
What the company is doing
On July 25, 2019, Neurological Corporation a subsidiary of Samsung Electronics issued Urgent Field Safety Notices to customers. Customers were advised to the following actions: 1) If the user wants to decrease the Dopper SV (Sample Volume) size set by himself/herself in the Doppler Only Mode, the user should press "set" button to update the PW Doppler mode. 2) A copy of the Field Safety Notice should be passed on to all who need to be aware within your organization. 3) Contact your local Samsung representative if you have questions. 4) Return the completed Customer Reply Form to Tom Leinart [email protected].
Recall details
- Recall number
- Z-2275-2019
- Status
- Terminated
- Initiated by the company
- Jul 25, 2019
- Recalling firm
- NeuroLogica Corporation
- Quantity
- 3582 worldwide
- Product code
- IYN
- FDA submissions
- K173713
- Distribution
- US Nationwide
Product description
HS70A Diagnostic Ultrasound System Version 2.01.00, 2.01.01, 2.01.02, 2.01.03, 2.01.04, 2.01.05, 2.01.06, 1.00.06, 1.00.07, 1.00.08, 1.00.09, 1.00.10, 1.00.11, 1.00.12
Code information (models, lots, serial numbers)
UDI: 08806167789220 Serial Numbers: S1C0M3HK300004A S1C0M3HK300005P S1C0M3HK300006D S1C0M3HK300007X S1C0M3HK300008N S1C0M3HK300001K S1C0M3HK300002Z S1C0M3HK400005Z S1C0M3HK400006Y S1C0M3HK400008P S1C0M3HK400001L S1C0M3HK400002T S1C0M3HK400003B S15NM3HK200005N S15NM3HK100001X S1C0M3HKC00001H S1C0M3HKC00002J S1C0M3HK900001Z S1C0M3HK900002Y S1C0M3HK900003A S1C0M3HK900004P S1C0M3HK900005D S1C0M3HK900006X S1C0M3HK900007N S1C0M3HK900008E S1C0M3HK900009F S1C0M3HK900010L S1C0M3HK900011T S1C0M3HK800001P S1C0M3HK800002D S1C0M3HK700001N S1C0M3HK700002E
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- HS70A Diagnostic Ultrasound System USS-HS7AL4L/WWSamsung Medison CO., Ltd. · Transducer, Ultrasonic, Diagnostic2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 83429).
- Z-2276-2019 · TerminatedWS80A Diagnostic Ultrasound System Version 1.00.16, 3.00.17, 3.00.18, 3.00.19, 3.01.00, 3.01.01, 3.01.02, 3.01.03, 3.01.04, 3.01.05, 3.01.06, 3.01.07, 3.01.08, 3.01.09, 4.00.08, 4.…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page