Alinity recallZ-2552-2019 · Sentinel CH SpA · initiated Jun 28, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Sentinel CH SpA or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Added an additional SmartWash parameter for the Lithium and Alinity c Lithium Reagent Kit, catalogue number 8L25-30 and 08P5320 to prevent the potential for carryover from the ARCHITECT Lactate Dehydrogenase (LDH) reagent (LN 2P56) may result infalsely elevated lithium patient results

FDA root cause category: Under Investigation by firm

What the company is doing

Sentinel issued Urgent Field Safety Notice letter on 6/28/19 via email to the distributor. The Distributor Abbott notified customers by letter Urgent Field Safety Notice Product Correction dated July 11, 2019 advising of the problem, health risk and action to take: If the LDH reagent is installed on your instrument, please follow the instructions 1. Update the Lithium (LN 8L25-30) assay parameters to add the new SmartWash. For additional information, see the ARCHITECT System Operations Manual, Section 2, Configure the SmartWash settings (c System. Install the Alinit y c Lithium assay file version 6 to implement the new LDH SmartWash. Customer Service at 18774ABBOTT (available 24 hours a day, 7 days a week). Customers outside the U.S., please contact your local area Customer Service

Recall details

Recall number
Z-2552-2019
Status
Terminated
Initiated by the company
Jun 28, 2019
Recalling firm
Sentinel CH SpA
Quantity
1217 kits
Product code
NDW
FDA submissions
K070987
Distribution
Nationwide Foreign: ARGENTINA AUSTRALIA AUSTRIA AZERBAIJAN BANGLADESH BELARUS BELGIUM BOSNIA AND HERZEGOVI BRAZIL BRUNEI CANADA CHILE COLOMBIA COSTA RICA CROATIA CZECH REPUBLIC DENMARK FAROE ISLANDS FINLAND FRANCE GERMANY GREECE GREENLAND HONG KONG IRELAND ISRAEL ITALY JORDAN KUWAIT LEBANON LUXEMBOURG MALAYSIA MALDIVES MEXICO MOROCCO NETHERLANDS NEW ZEALAND NIGERIA NORWAY OMAN PAKISTAN PANAMA PARAGUAY PHILIPPINES POLAND PORTUGAL PUERTO RICO ROMANIA RWANDA SAUDI ARABIA SLOVAKIA SOUTH AFRICA SPAIN ST. VINCENT SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UAE UNITED KINGDOM URUGUAY

Product description

Alinity c Lithium Reagent Kit- intended for the quantitation of lithium in serum or plasma using respectively the ARCHITECT c Systems and the Alinity c analyzer List Number: 08P5320

Code information (models, lots, serial numbers)

Lot Number/UDI: 80553Y600/ (01)00380740136277 (17)190731(10)80553Y600; 80444Y600/ (01)00380740136277 (17)190731(10)80444Y600; 80625Y600/ (01)00380740136277 (17)191130(10)80625Y600; 90224Y600/ (01)0038074013627 (17)200229(10)90224Y600