Sapphire recallZ-0129-2020 · ICU Medical Inc · initiated Jul 29, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from ICU Medical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for fluid leakage out of air vents on inline filters on Plum and Sapphire micropore infusion sets.
FDA root cause category: Under Investigation by firm
What the company is doing
ICU Medical notified customers, via Stericycle, Inc., on about 07/29/2019 via "URGENT: MEDICAL DEVICE CORRECTION" letter. Customers were instructed to check inventory and quarantine any sets that are affected by the recall, return them to ICU Medical, inform potential users within your organization of the recall, and complete and return the provided response form. Please notify customers if affected product was further distributed and request that they contact Stericycle at 1-855-311-5437 (M-F, 8am-5pm ET) to receive a reply form. Customers were also requested to consider the following situations: 1. In applications where the use of a filter is not deemed necessary, consider utilizing an administration set without a filter. 2. In applications where filter use is desired and microbore tubing is not required, consider using filtered macrobore sets in place of microbore sets. 3. In cases where a filtered macrobore set is desired but no alternative macrobore set is available contact ICU Medical because a non-filtered set along with an add on filter or filtered extension set could be considered. 4. In instances where the benefits of a filtered microbore infusion set are greater than the potential risks of a filter leak, the product affected by the recall may be utilized out of medical necessity. In these cases, an individualized assessment of benefits and risks should be undertaken with consideration of the unique characteristics of the patient and the infusion itself. When the benefits exceed the risks and the set is deemed medically necessary, users should periodically monitor the vented filter area for leakage. If leaking is observed, consider changing to a new infusion set as soon as clinically appropriate. Report this event to ICU Medical.
Recall details
- Recall number
- Z-0129-2020
- Status
- Terminated
- Initiated by the company
- Jul 29, 2019
- Recalling firm
- ICU Medical Inc
- Quantity
- 73,710 units
- Product code
- FRN
- FDA submissions
- K171346
- Distribution
- Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY. International distribution to Canada, Chile, France, Germany, Greece, Hong Kong, Ireland, Italy, Oman, Philippines, Poland, Saudi Arabia, South Africa, Spain, Taiwan, Turkey, United Arab Emirates, United Kingdom.
Product description
Infusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Primary 1.2 Micron Filter Set Microbore, 122 Inch. 1 unit per pouch; 50 pouches per case
Code information (models, lots, serial numbers)
Product Code: 163350455, Lot number (UDI Number): 855935H (20887787007316, 10887787007319), 855965H (20887787007316, 10887787007319), 876145H (20887787007316, 10887787007319), 885395H (20887787007316, 10887787007319), 896135H (20887787007316, 10887787007319), 936605H (20887787007316, 10887787007319), 3843263 (20887787007316, 10887787007319), 3864786 (20887787007316, 10887787007319)
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Sapphire 1633503ICU Medical, Inc. · ACCESSORIES, PUMP, INFUSION1 recall0 reports
Product lines
Other devices under the same FDA submission
9 devices cleared under K171346 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Sapphire 1636202ICU Medical, Inc. · ACCESSORIES, PUMP, INFUSION1 recall3 reports
- Sapphire 1638401ICU Medical, Inc. · ACCESSORIES, PUMP, INFUSION1 recall1 report
- Sapphire 1636302ICU Medical, Inc. · ACCESSORIES, PUMP, INFUSION1 recall0 reports
- Sapphire 1637302ICU Medical, Inc. · ACCESSORIES, PUMP, INFUSION1 recall0 reports
- Sapphire 1638303ICU Medical, Inc. · ACCESSORIES, PUMP, INFUSION1 recall0 reports
- Sapphire 2401001ICU Medical, Inc. · ACCESSORIES, PUMP, INFUSION1 recall0 reports
- Sapphire 2401301ICU Medical, Inc. · ACCESSORIES, PUMP, INFUSION1 recall0 reports
- Sapphire 1632402ICU Medical, Inc. · ACCESSORIES, PUMP, INFUSION1 report
- Sapphire 1637402ICU Medical, Inc. · ACCESSORIES, PUMP, INFUSION1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 83549).
- Z-0126-2020 · TerminatedInfusion Pump I.V. Administration, specifically Plum Microbore Infusion Sets with Inline Filters labeled as High-Pressure Filter PRIMARY I.V. PLUMSET Convetible Pin, 104 Inch with …
- Z-0127-2020 · TerminatedInfusion Pump I.V. Administration, specifically Plum Microbore Infusion Sets with Inline Filters labeled as Primary PLUM Set 0.2 Micron Filter, Prepierced Y-Site, Polyethylene Line…
- Z-0128-2020 · TerminatedInfusion Pump I.V. Administration, specifically Plum Microbore Infusion Sets with Inline Filters labeled as Primary PLUM Set Yellow Striped Tubing, Distal Microbore Tubing, 0.2 Mic…
- Z-0130-2020 · TerminatedInfusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Primary 0.2 Micron Filter Set, Microbore, CLAVE Y-Site, 118…
- Z-0131-2020 · TerminatedInfusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Epidural Set Yellow-Striped Microbore, 0.2 Micron Filter 11…
- Z-0132-2020 · TerminatedInfusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Epidural Set Yellow-Striped Microbore, 0.2 Micron Filter, P…
- Z-0133-2020 · TerminatedInfusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Primary 0.2 Micron Filter Set Non-Vented, Microbore, PAV, C…
- Z-0134-2020 · TerminatedInfusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Primary 1.2 Micron Filter Set Non-Vented, Microbore, PAV, 1…
- Z-0135-2020 · TerminatedInfusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Epidural Set with NRFit Connector Yellow-Striped Microbore,…
- Z-0145-2020 · TerminatedInfusion Pump I.V. Administration, specifically Plum Microbore Infusion Sets with Inline Filters labeled as PLUM Epidural Set with NRFit Connector Yellow-Striped Tubing, Distal Mic…
- Z-0147-2020 · TerminatedInfusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Epidural Set with NRFit Connector Yellow-Striped Microbore,…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page