Origin Data Management / Patient Data Manager recallZ-0176-2020 · Brainlab AG · initiated Sep 25, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Brainlab AG or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The Brainlab Patient Data Manager software (PDM) in combination with the Brainlab iPlan CMF/Cranial/ENT/Spine software 2.1.1, 2.2.0, 2.2.1, 2.3.0, 2.3.1, 2.4.0, 2.5.0, 2.5.1 (if used in combination with iPlan CMF/Cranial/ENT/Spine 3.0.5, 3.0.6) that may lead to shifts applied to segmentation objects in iPlan being not reflected when exported into DICOM under specific circumstances.
FDA root cause category: Software design
What the company is doing
As of September 25, 2019 for new installations or updates/upgrades to the affected SW version, the Field Safety Notice / Product Notification letter is hand delivered to the customer during installation (before clinical use) by the according Brainlab Service representative All other consignees were notified by email on 10/04/2019. The letter details how the even occurs and provides instruction on how to avoid the issue until a software revision is disseminated.
Recall details
- Recall number
- Z-0176-2020
- Status
- Terminated
- Initiated by the company
- Sep 25, 2019
- Recalling firm
- Brainlab AG
- Quantity
- 1230 units
- Product code
- LMA
- Distribution
- worldwide
Product description
Patient Data Manager, Model/Catalogue Numbers: 25100-09 - PATIENT DATA MANAGER 2.1.1LICENSE ONLY 25100-10 - PATIENT DATA MANAGER 2.2.0 LICENSE ONLY 25100-11 - PATIENT DATA MANAGER 2.2.1 LICENSE ONLY 25100-12 - PATIENT DATA MANAGER 2.3 INSTALLER 25100-12A - PATIENT DATA MANAGER 2.3.1 INSTALLER 25100-13 - PATIENT DATA MANAGER 2.3 LICENSE 25100-12B - PATIENT DATA MANAGER 2.4.0 INSTALLER 25100-13A - PATIENT DATA MANAGER 2.4 LICENSE 25100-12C - PATIENT DATA MANAGER 2.5.0 INSTALLER 25100-12D - PATIENT DATA MANAGER 2.5.1 INSTALLER 25100-13B - PATIENT DATA MANAGER 2.5 LICENSE used in combination with iPlan CMF/Cranial/ENT/Spine 3.0.5, 3.0.6. Product Usage: Patient Data Manager is intended to load patient data from and export patient data to connected network locations (like file shares and PACS) and removable media. Furthermore, the software is intended to start other applications (like Brainlab planning and navigation applications) and to route applications and connected video sources to displays attached to the system that are running Patient Data Manager. The software is also intended to connect two Brainlab systems to each other via network.
Code information (models, lots, serial numbers)
Patient Data Manager versions 2.1.1 no GTIN 2.2.0 no GTIN 2.2.1 no GTIN 2.3.0, GTIN 04056481006662 2.3.1, GTIN 04056481132453 2.4.0, GTIN 04056481138615 2.5.0, GTIN 04056481140502 2.5.1, GTIN 04056481140502 versions 2.1.1, 2.2.0, 2.2.1, 2.3.0, 2.3.1, 2.4.0, 2.5.0, 2.5.1.
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- Origin Data Management / Patient Data Manager 2.3Brainlab SE · Digitizer, Image, Radiological1 recall0 reports
- Origin Data Management / Patient Data Manager 2.3.1Brainlab SE · Digitizer, Image, Radiological1 recall0 reports
- PATIENT DATA MANAGER 2.4 INSTALLERBrainlab SE · Digitizer, Image, Radiological1 recall0 reports
- Patient Data Manager 2.5Brainlab SE · Digitizer, Image, Radiological1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page