Skip to content

Giraffe OmniBed Carestation, Model Numbers:… recallZ-0539-2020 · GE Healthcare, LLC · initiated Oct 11, 2019

Class I Terminated All recalls posted in December 2019

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.

FDA root cause category: Device Design

What the company is doing

The firm disseminated its notices by letter beginning on 10/11/2019. The notices identified the following safety issues: - The bedside panels of Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation can be upright and look closed but not be latched. - The portholes also can look closed when not latched. - If a canopy cover is used, it can hold the bedside panel or porthole door closed without being latched. The product may continue to be used following specific safety instructions outlined in the notice. On/about 11/8/2019, the firm issued an "Amended - Urgent Medical Device Recall" letter that supplemented the previous notification to provide additional safety instructions and materials (poster regarding risk of patient fall, user manual addendum, and labeling, including safety labels). Update 6/3/2021: In the event a new consignee was identified, GE Healthcare has sent the original letter.

Recall details

Recall number
Z-0539-2020
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
Oct 11, 2019
Posted by the FDA
Dec 3, 2019
Recalling firm
GE Healthcare, LLC
Quantity
7814 units
Product code
FMT
FDA submissions
K152814
Distribution
worldwide

Product description

Giraffe OmniBed Carestation, Model Numbers: 2082844-001-XXXXXX; Giraffe OmniBed Carestation CS1, Model Numbers: 2082844-001-XXXXXXXX. infant warmer

Code information (models, lots, serial numbers)

Giraffe OmniBed Carestation, Model Numbers: 2082844-001-XXXXXX Giraffe OmniBed Carestation CS1, Model Numbers: 2082844-001-XXXXXXXX Updated 6/3/2021 - All model numbers and units are affected.

Other devices under the same FDA submission

1 device cleared under K152814 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 84077).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page