Medivators EndoGator REF 100651. Kit for Olympus… recallZ-1777-2020 · Medivators, Inc. · initiated Oct 9, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medivators, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medivators is recalling Endo SmartCap Tubing due to a manufacturing anomaly that may present during procedures as inadequate insufflation, presence of air bubbles inside the endo smartcap tubing, insufficient water to rinse the endoscope lens, or water leaking from the air/water valve.
FDA root cause category: Equipment maintenance
What the company is doing
An Urgent Medical Device Recall Notification letter was sent to customers via certified mail next day delivery on 10/09/2019. The letter identified affected product, stated reason for recall, asked for product to be quarantined and discontinue use. Customers are to return product to Medivators by obtaining a Return Material Authorization from Customer Service. A response form was asked to be returned. Questions can be directed to Customer Service at 1-800-444-4729.
Recall details
- Recall number
- Z-1777-2020
- Status
- Terminated
- Initiated by the company
- Oct 9, 2019
- Recalling firm
- Medivators, Inc.
- Quantity
- 53,790 units total
- Product code
- FAJ
- FDA submissions
- K093665
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Australia, England.
Product description
Medivators EndoGator REF 100651. Kit for Olympus 140/160/180/190 Series GI Endoscopes for CO2 Connection with Endogator EGP-100, Olympus OFP, Olympus AFU-100 & Erbe EIP 2 Irrigation Pump Units. Contents 10. Sterile EO. of 100145CO2EXT Endo SmartCap Irrigation Tubing for Olympus 140/160/180/190 Series GI Endoscopes with Extended CO2 Input - Product Usage: is intended to be used with an air or CO2 source and/or a pump along with a sterile water source to supply air or CO2 and sterile water to an endoscope during endoscopic procedures.
Code information (models, lots, serial numbers)
Lot # / Case UDI#s: 430358 / 2067796406530322060110430358, 429884 / 2067796406530322060110429884, 428806 / 2067796406530322070110428806, 430568 / 2067796406530322060110430568, 431254 / 2067796406530322070110431254, 431455 / 2067796406530322070110431455, 431603 / 2067796406530322070110431603, 431834 / 2067796406530322070110431834, 431985 / 2067796406530322070110431985, 431973 / 2067796406530322070110431973, 432922 / 2067796406530322070110432922, 432108 / 2067796406530322070110432108, 432921 / 2067796406530322070110432921, 433116 / 2067796406530322070110433116, 433115 / 2067796406530322080110433115.
Other devices under the same FDA submission
5 devices cleared under K093665 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ENDO SMARTCAP 100145CO2EXTMedivators Inc. · Cystoscope And Accessories, Flexible/Rigid21 reports
- Endo SmartCap 100164CO2EXTMedivators Inc. · Cystoscope And Accessories, Flexible/Rigid4 reports
- ENDO SMARTCAP 100165CO2EXTMedivators Inc. · Cystoscope And Accessories, Flexible/Rigid2 reports
- ENDO SMARTCAP 100160Medivators Inc. · Cystoscope And Accessories, Flexible/Rigid2 reports
- ENDO SMARTCAP 100150CO2Medivators Inc. · Cystoscope And Accessories, Flexible/Rigid1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 84109).
- Z-1772-2020 · TerminatedMedivators Endo Smartcap Endoscope Tubing. REF100145. Packaged as Contents 1, 10, 50, and found in BMP-C145. Tubing for Olympus 140/240, 160/260, 180/280, 190/290 Series Endosco…
- Z-1773-2020 · TerminatedMedivators Endo Smartcap Endoscope Tubing. REF100145CO2, packaged as Contents 1, 10, 50 and found in BMP-C145CO2. Tubing for Olympus 140/240, 160/260, 180/280, 190/290 Series E…
- Z-1774-2020 · TerminatedMedivators Endo Smartcap Endoscope Tubing. REF100145CO2EXTU, packaged as Contents 1, 10, 50. Tubing for Olympus 140/240, 160/260, 180/280, 190/290 Series Endoscopes With Extend…
- Z-1775-2020 · TerminatedMedivators Endo Smartcap Endoscope Tubing. REF100145CO2EXT, packaged as Contents 1, 10, 50, and found in BMP-C145CO2EXT. Tubing for Olympus 140/240, 160/260, 180/280, 190/290 S…
- Z-1776-2020 · TerminatedMedivators EndoGator REF 100650. Kit for Olympus 140/160/180/190 Series GI Endoscopes forCO2 Connection with Olympus OFP or Boston Scientific Endostat II Irrigation Pump Units. C…
- Z-1778-2020 · TerminatedMedivators Endo carry-on Procedure Kit REF CPK-100222-0002 - Product Usage: is intended to be used with an air or CO2 source and/or a pump along with a sterile water source to sup…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page