Torcon NB recallZ-0435-2020 · Cook Inc. · initiated Oct 21, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cook Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches of Torcon NB Advantage Catheters.
FDA root cause category: Process change control
What the company is doing
On October 21, 2019 the firm distributed Urgent Medical Device Recall Communication letters by mail instructing them to: 1. Examine inventory immediately to determine if you have affected product(s), and quarantine affected product(s). Immediately cease all distribution and use of this product. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Note: Unaffected products that are returned will not be credited. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email ([email protected]). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: [email protected]. Adverse events or quality problems experienced with the use of this product may also be reported to the FDA: " Visit http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm to obtain a form to fax or mail. " Call the FDA at 800.FDA.1088. We recognize that this situation is a potential disruption to your normal operations, and we sincerely apologize. Thank you for your immediate assistance in this matter. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235. We look forward to your response.
Recall details
- Recall number
- Z-0435-2020
- Status
- Terminated
- Initiated by the company
- Oct 21, 2019
- Recalling firm
- Cook Inc.
- Quantity
- 30
- Product code
- DQO
- FDA submissions
- K133130
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, IL, IN, KS, LA, MA, MD, Ml, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WV, and PR. Countries of AU, BE, CA, CN, CZ, DK, FR, DE, GT, HK, IT, KR, LU, MX, NL, PK, PL, ZA, ES, SE, CH, TW, and GB.
Product description
Torcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicians trained and experienced in angiographic techniques.
Code information (models, lots, serial numbers)
REF (RPN): HNB5.0-38-65-P-NS-RIM REF (GPN): G05979 Lot Number: 9890936 UDI: (01)00827002059795(17)220729(10)9890936
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Torcon NB G05979Cook Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 84125).
- Z-0421-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0422-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0423-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0424-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0425-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0426-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0427-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0428-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0429-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0430-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0431-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0432-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0433-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0434-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0436-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0437-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0438-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0439-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0440-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0441-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0442-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0443-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0444-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0445-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0446-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0447-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0448-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0449-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0450-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
- Z-0451-2020 · TerminatedTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicia…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page