Skip to content

Intramedullary Bone Saw recallZ-0594-2020 · Zimmer Biomet, Inc. · initiated Oct 25, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zimmer Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Possible commingle of 13 MM and 14 MM Intramedullary Bone Saw Cam Assembly -an Intramedullary Bone Saw Cam Assembly was labeled and etched as a 14 MM but measured as a 13 MM cam

FDA root cause category: Error in labeling

What the company is doing

Zimmer Biomet issued URGENT MEDICAL DEVICE RECALL letter via FedEx and email on 10/25/19. The letter identifies the reason for recall, health risk and action to take: Review this notification and ensure that affected personnel are aware of the contents. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. Questions call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Alternatively, your questions may be emailed to [email protected].

Recall details

Recall number
Z-0594-2020
Status
Terminated
Initiated by the company
Oct 25, 2019
Recalling firm
Zimmer Biomet, Inc.
Quantity
26 units
Product code
LXH
Distribution
Distribution to US states of CA, FL, IN, MN, MO, NY, and UT, Australia, and New Zealand

Product description

Biomet Orthopedics Intramedullary Bone Saw Cam Assembly 13MM-Orthopedic manual surgical instrument Item Number: 475660

Code information (models, lots, serial numbers)

Lot Numbers: 025800, 025810, 081640 UDI:  (01)00880304386570(17)280202(10)025800 (01)00880304386570(17)280205(10)025810 (01)00880304386570(17)280524(10)081640

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 84224).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page