HEALIX ADVANCE recallZ-0666-2020 · DePuy Mitek, Inc., a Johnson & Johnson Co. · initiated Nov 4, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from DePuy Mitek, Inc., a Johnson & Johnson Co. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Specific product codes and lots of the Mitek HEALIX ADVANCE" Knotless Anchors were manufactured with BIOCRYL" Biocomposite Material instead of BIOCRYL RAPIDE Biocomposite Material. BIOCRYL and BIOCRYL RAPIDE are biocompatible and effective material choices for absorbable, injection-molded sports medicine implant fixation.
FDA root cause category: Employee error
What the company is doing
On November 4, 2019 the firm distributed Urgent Medical Device Recall Communication letters by mail: URGENT FIELD SAFETY NOTICE PRODUCT RECALL HEALIX ADVANCE" Knotless Anchors Dear Valued Customer, On 04-November-2019, DePuy Mitek, Inc. initiated a Voluntary Product Removal of specific product codes and lots of the Mitek HEALIX ADVANCE" Knotless Anchors. The anchors were manufactured with BIOCRYL" Biocomposite Material instead of BIOCRYL RAPIDE Biocomposite Material. BIOCRYL and BIOCRYL RAPIDE are biocompatible and effective material choices for absorbable, injection-molded sports medicine implant fixation. No complaints or adverse events related to this issue have been received for the affected product codes/lots. Product Code: 222330 Product Description: HEALIX KNOTLESS ADV BR 4.75 GTIN: 10886705023462 Lot Number: 5L45257 Product Code: 222331 Product Description: HEALIX KNOTLESS ADV BR 5.5 GTIN: 10886705023479 Lot Numbers: 5L13543 and 5L45259 Potential Harms: Based on the analysis, it is expected that sport medicine implants made from BIOCRYL (PLLA+TCP) would perform equal to those made from BIOCRYL RAPIDE (PLGA + TCP) during the surgery and therapeutic healing period. As such, this product issue presents no foreseeable increased risk to patient safety nor does it alter the product benefit-risk profile. BIOCRYL and BIOCRYL RAPIDE Comparison: " Each material is used clinically today, but in different DePuy Mitek implant brands; both materials have a more than 15 years of use in clinical applications. BIOCRYL was first launched in 2002 and BIOCRYL RAPIDE in 2004. " The mechanical properties (strength, stiffness) of BIOCRYL and BIOCRYL RAPIDE are similar at the time of anchor deployment; BIOCRYL offers equivalent in vitro strength retention during the therapeutic healing period and out to 12 weeks. " As for long-term absorption rates, BIOCRYL RAPIDE is expected to absorb in approximately two years versus approximately four years
Recall details
- Recall number
- Z-0666-2020
- Status
- Terminated
- Initiated by the company
- Nov 4, 2019
- Recalling firm
- DePuy Mitek, Inc., a Johnson & Johnson Co.
- Quantity
- 599
- Product code
- MAI
- FDA submissions
- K130917
- Distribution
- US: AL AZ CA CT FL IA IL IN LA MA MD ME MI MN MT NC NE NH NJ NM OH OK OR PA TN TX VA WA WI International: Belgium, Chile, China, Czech Republic, France, India, Japan, Korea, Latvia, New Zealand, Slovakia Sweden
Product description
Healix Knotless ADV BR 5.5 Suture Anchor
Code information (models, lots, serial numbers)
Part Number: 222331 Batch/Lot: 5L45259, 5L13543 GTIN: 10886705023479
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall196 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K130917 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall607 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 84237).
- Z-0665-2020 · TerminatedHealix Knotless ADV BR 4.75 Suture Anchor
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.