Omnifit recallZ-0811-2020 · Howmedica Osteonics Corp. · initiated Nov 7, 2019
Class II Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Howmedica Osteonics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
May have a missing locking wire causing delay in surgery while retrieving a back-up device or implantation of a device that is missing a locking wire
FDA root cause category: Employee error
What the company is doing
Stryker issued Urgent Medical Device Recall - PFA 2147264 letter dated November 7, 2019) to US consignees by third-party recall company, Stericycle Inc. (2670 Executive Drive Indianapolis, IN 46241), via UPS 2nd day air. Letter states reason for recall, health risk and action to take: Actions Needed: 1.Please inform users of this Urgent Medical Device Recall and forward this notice to all those individuals who need to be aware within your organization. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. A response is required, even though you may not have any physical inventory on site anymore. 3.Quarantine and discontinue use of the recalled devices. 4.Hospitals/Branches/Agencies: Complete and sign the enclosed Urgent Medical Device Recall Business Reply Form and fax a copy to 1-888-350-3903 or email to [email protected] within 5 Days. 5.Hospitals Only: Return all affected product available at your location to your local branch office. 6.Branches/Agencies: Return all affected product returned by the hospital and/or available at your location to the following address: Stryker Orthopaedics/PFA Product Returns, Attn: Distribution Inventory Team 325 Corporate Drive, Dock M-East, Mahwah, NJ 07430, PFA #2147264. Actions Needed: 1.Please inform users of this Urgent Medical Device Recall and forward this notice to all those individuals who need to be aware within your organization.2.Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. A response is required, even though you may not have any physical inventory on site anymore. 3.Quarantine and discontinue use of the recalled devices. 4.Hospitals/Branches/Agencies: Complete and sign the enclosed Urgent Medical Device Recall Business Reply Form and fax a copy to
Recall details
- Recall number
- Z-0811-2020
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 7, 2019
- Recalling firm
- Howmedica Osteonics Corp.
- Quantity
- 20 units
- Product code
- JDI
- FDA submissions
- K974685
- Distribution
- CA, FL, GA , IA, MA, MI, NC, OH, WI Foreign: Canada, Japan
Product description
Stryker Omnifit Crossfire 10 Degree Series II Insert- intended for cementless and/or cemented fixation within the prepared acetabulum Catalog #: 2041C-3254
Code information (models, lots, serial numbers)
Lot Number: 53778501 UDIs: (01)04546540035806(17)201108(10)53778501
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Omnifit 2041C-3254Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall2 reports
Product lines
Other devices under the same FDA submission
12 devices cleared under K974685 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Omnifit 2041C-2846Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented5 reports
- Omnifit 2041C-3250Howmedica Osteonics Corp. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 reports
- Omnifit S-2301C-2846Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented2 reports
- Omnifit 2041C-2658Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented1 report
- Omnifit 2041C-2858Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 report
- Omnifit 2041C-3258Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented1 report
- Omnifit 2043C-2846Howmedica Osteonics Corp. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- Omnifit 2043C-2850Howmedica Osteonics Corp. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- Omnifit 2043C-2854Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented1 report
- Omnifit S-2301C-3262Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 report
- Omnifit S-2302C-3250Howmedica Osteonics Corp. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- Omnifit 2041C-2246Howmedica Osteonics Corp. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page