ADVIA Chemistry CardioPhase High Sensitivity C-Reactive… recallZ-1260-2020 · Siemens Healthcare Diagnostics, Inc. · initiated Nov 19, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Atellica and ADVIA hsCRP calibrator accuracy has drifted away from the reference materials (ERM-DA470/IFCC), which is causing a high bias on QC and patient sample recovery.

FDA root cause category: No Marketing Application

What the company is doing

On November 19, 2019 the firm distributed Urgent Medical Device Recall Communication letters by mail stating: Urgent Medical Device Correction Our records indicate that your facility may have received the following product: Atellica CH 930 Analyzer Reassignment of the Atellica CH High Sensitivity C-Reactive Protein (hsCRP) Calibrator Lot 453716 (Ref/Siemens Material Number 11099412/11099412) ADVIA Chemistry Systems Reassignment of the ADVIA Chemistry CardioPhase High Sensitivity C-Reactive Protein (hsCRP) Calibrator Lot 453610 (Ref/Siemens Material Number -05006455/10335897) Reason for Correction The purpose of this communication is to inform you of an issue with the product indicated above and provide instructions on actions that your laboratory must take. Siemens has confirmed that the current assigned values for ADVIA Chemistry hsCRP Calibrator Lot 453610 demonstrate a positive bias of approximately 11% for hsCRP patient samples and Quality Control material when compared to the European Reference Material ERM DA474/IFCC for CRP measurement. Depending on the QC concentrations and ranges used by the laboratory, QC may not detect this issue. To correct for the positive bias, the values for ADVIA Chemistry hsCRP Calibrator Lot 453610 have been reassigned to align with reference material ERM DA474/IFCC. The information in Table 2 supersedes the information in the calibrator lot-specific value sheet for hsCRP Calibrator Lot 453610. Refer to Table 2 for the reassigned calibrator values. Patient and Quality Control results are expected to shift approximately -11% when using the reassigned calibrator values for Lot 453610. Based on the negative shift in recovery, it may be necessary to adjust your laboratorys quality control ranges. Refer to Table 3 for representative quality control recovery data. Future commercial calibrator lots will be traceable to this new reference material and will demonstrate comparable performance to the reassig

Recall details

Recall number
Z-1260-2020
Status
Terminated
Initiated by the company
Nov 19, 2019
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
844
Product code
JIX
FDA submissions
K081294
Distribution
Worldwide distribution US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, WA, WY and countries of UK, CA, PH, MX, VN, TW, TH, IN, CO, AE, AR, AU, BD, BE, BG, BR, CL, CZ, DE, DK, EG, ES, FI, FR, GR, HU, IE, IT, KR, MY, NL, PL, PT, RO, RU, SE, SG.

Product description

ADVIA Chemistry CardioPhase High Sensitivity C-Reactive Protein (hsCRP) Calibrator - product Usage: For in vitro diagnostic use in the calibration of ADVIA Chemistry systems for the CardioPhase hsCRP assay.

Code information (models, lots, serial numbers)

Siemens Material Number (SMN): ADVIA Chemistry: 10335897 REF: 05006455 Lot# 453610 Expiration Date: 2020-01 UDI Number: ADVIA: (01)00630414526843(10)453610(17)20200128

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K081294 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 84413).