Dexcom G6 recallZ-1104-2020 · Dexcom Inc · initiated Nov 19, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Dexcom Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
It was reported that the user's low alarm feature on the iOS application were not properly alerting users when the user has enabled the Alert Schedule feature more than 30 days after installing the iOS application.
FDA root cause category: Software design
What the company is doing
An updated G6 iOS CGM App was placed in the Apple App Store (SW11677 version 1.4.4) with messaging pushed to all US users on November 19, 2019. The messaging repeats to the user on a daily basis until the user updates (or is alternatively blocked from use of the app if they ignore the messaging and do not update). Messaging will be delivered directly to the end user through the following App Compatibility Messaging via the Apple App Store (see Compatibility Message and time-frame below): Duration: Day 1-3 Date Range: Nov 19-21 App Compatibility Message: Warning - Please update your Dexcom G6 iOS App from the Apple App Store. Failure to update your App may cause your alarms and alerts to not work as intended. Without alarms/alerts, you might miss a severe low or high glucose event. Contact technical support if you require assistance. Duration: Day 4-17 Date Range: Nov 22 - Dec 5 App Compatibility Message: Warning - This version of the Dexcom G6 iOS App will no longer be available for use as of December 6. Update the Dexcom G6 iOS App now from the Apple App Store to continue use of the Dexcom G6 iOS App. Contact technical support if you require assistance. Duration: Day 18 Date Range: Dec 6 App Compatibility Message: Warning - This version of the Dexcom G6 iOS App is no longer available for use. Update the Dexcom G6 iOS App now from the Apple App Store to continue use of the Dexcom G6 iOS App. Contact technical support if you require assistance. For impacted users, they will receive the Alert Settings Reset notification once the app update is complete and the app is launched or the app receives a communication from the Transmitter. The user must then reset any alarm/alert customizations made (or leave at factory default settings). For users currently not impacted, they will not receive the Alert Settings Reset notification. The updated version of the app corrects the problem for those impacted and prevents the problem from occurring for
Recall details
- Recall number
- Z-1104-2020
- Status
- Terminated
- Initiated by the company
- Nov 19, 2019
- Recalling firm
- Dexcom Inc
- Quantity
- 237888 users registered.
- Product code
- QBJ
- FDA submissions
- K191450
- Distribution
- US Nation-wide OUS: Austria; Belgium; Canada; Czechia; Denmark; Finland; Germany; Great Britain; Hong Kong; Ireland; Italy; Luxembourg; Netherlands; Norway; South Africa; Spain; Sweden, and Switzerland.
Product description
Brand Name: Dexcom G6 CGM App for iOS Software Number: SW11677 Software Version: Version 1.4.0 up to corrected Version 1.4.4 GUDID Registered GTIN: 00386270000804
Code information (models, lots, serial numbers)
Software Number: SW11677 Software Version: 1.4.0 up to corrected Version 1.4.4 GUDID Registered GTIN: 00386270000804
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Dexcom G6 iOS SW11677Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall23,526 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page