O-ARM recallZ-0767-2020 · Medtronic Navigation, Inc.-Littleton · initiated Nov 15, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Navigation, Inc.-Littleton or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to a manufacturing issue affecting particular O-arm" O2 Imaging Systems, Medtronic has determined that the gantry tractor motor drive belt may loosen over an extended number of uses at a higher frequency in these systems. If the belt loosens, it may result in rotation of 3D images about the gantry isocenter. While the 3D image is anatomically accurate within the image itself and may be used to confirm therapy, its electronic registered location may be rotated relative to the actual physical position of the patient.
FDA root cause category: Device Design
What the company is doing
On November 15, 2019 the firm distributed Urgent Medical Device Recall Communication letters by mail stating: MEDICAL DEVICE CORRECTION O-arm" O2 Imaging System Dear Healthcare Professional: Medtronic has determined through internal testing that there is a potential for navigational inaccuracy when utilizing the O-armTM O2 Imaging Systems auto-registration feature, when used in conjunction with an Image-Guided Surgery System for a specific set of O-armTM O2 Imaging System serial numbers. A list of potentially affected systems is provided in Table 1 of this letter. Due to loosening of an internal motor drive belt over extended number of uses, image rotation may occur relative to patient positioning data. Inaccuracy may occur when the O-arm registration information is transferred to an Image-Guided Surgery System to be used in navigation, utilizing the rotated images. Navigational inaccuracies are inherent in the use of any Image-Guided Surgery System utilizing an O-armTM O2 Imaging System, but the potential for inaccuracy may be higher if image rotation occurs in an affected system. This letter provides awareness of this issue and reinforces existing instructions within the O-arm" O2 Imaging System User Manual (IFU) that allow for identification of the issue if it occurs. This issue does not involve the O-arm 1000 Imaging Systems. Issue Background and Summary: The O-arm" O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. Medtronic has determined that if the gantry tractor motor drive belt loosens over extended number of uses, it can result in rotation of 3D images about the gantry isocenter. While the 3D image is anatomically accurate within the image itself and may be used to confirm therapy, its electronic registered location may
Recall details
- Recall number
- Z-0767-2020
- Status
- Terminated
- Initiated by the company
- Nov 15, 2019
- Posted by the FDA
- Dec 31, 2019
- Recalling firm
- Medtronic Navigation, Inc.-Littleton
- Quantity
- 90
- Product code
- OWB
- FDA submissions
- K151000, K173664
- Distribution
- US: AR CA CO CT FL ID IL IN KS KY MI MN MO NC ND NE NY OH OR PA SC SD TX VA WA WI WV OUS: AUSTRALIA CANADA CHINA FRANCE GERMANY INDIA ITALY JAPAN LEBANON NETHERLANDS ROMANIA SAUDI ARABIA SPAIN SWEDEN SWITZERLAND UNITED KINGDOM
Product description
O-arm" O2 Imaging System, Product Number/CFN: BI-700-02000
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
SN: C2155 C2180 C2193 C2203 C2213 C2223 C2233 C2243 C2257 C2267 C2157 C2181 C2194 C2204 C2214 C2224 C2234 C2247 C2258 C2268 C2165 C2182 C2195 C2205 C2215 C2225 C2235 C2248 C2259 C2166 C2183 C2196 C2206 C2216 C2226 C2236 C2250 C2260 C2167 C2185 C2197 C2207 C2217 C2227 C2237 C2251 C2261 C2169 C2186 C2198 C2208 C2218 C2228 C2238 C2252 C2262 C2172 C2187 C2199 C2209 C2219 C2229 C2239 C2253 C2263 C2176 C2189 C2200 C2210 C2220 C2230 C2240 C2254 C2264 C2177 C2190 C2201 C2211 C2221 C2231 C2241 C2255 C2265 C2178 C2192 C2202 C2212 C2222 C2232 C2242 C2256 C2266 UDIs: 00763000074104, 00763000132378, 00763000074128, 00763000134488 00763000081003, 00763000074067, 00763000168605, 00763000074135, 00763000134518, 00763000081041, 00763000135829, 00763000081102, 00763000074050, 00763000074098, 00763000135836
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- O-ARM® BI70002000Medtronic Navigation, Inc. · Image-intensified fluoroscopic x-ray system, mobile2 recalls2,132 reports
- O-ARM® BI70002000Medtronic Navigation, Inc. · Image-intensified fluoroscopic x-ray system, mobile2 recalls65 reports
- O-ARM® BI70002000Medtronic Navigation, Inc. · Interventional fluoroscopic x-ray system1 recall20 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page