Skip to content

Fast-Cath recallZ-0789-2020 · Abbott Medical · initiated Dec 9, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Medical or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Three lots of Fast-Cath Hemostasis Introducer devices contained 8F sized dilators and sheaths instead of the expected 8.5F devices.

FDA root cause category: Packaging

What the company is doing

Beginning on December 9, 2019, US consignees will be notified in person by an Abbott field representative. OUS consignees will be notified by mail. The firm is seeking return of the devices.

Recall details

Recall number
Z-0789-2020
Status
Terminated
Initiated by the company
Dec 9, 2019
Recalling firm
Abbott Medical
Quantity
380 devices
Product code
DYB
FDA submissions
K914090
Distribution
Worldwide distribution - US Nationwide distribution in the states of MA, VA, NH, FL, CT, UT, OH, WI, NY, IA, MI, MS, PA, and countries of Canada, Czech Republic

Product description

Fast-Cath Hemostasis Introducer, HEMOSTASIS INTRODUCER, 8.5F, 2.83 mm, REF 406143 - Product Usage: The Fast-Cath Hemostasis Introducers are indicated for the introduction of angiographic catheters, closed-end catheters, balloon catheters, and electrodes into a vessel where minimizing blood loss is essential.

Code information (models, lots, serial numbers)

Batch 6917183; UDI (01)15414734014219(17)220228(10)6917183  Batch 6950628; UDI (01)15414734014219(17)220331(10)6950628  Batch 7099144; UDI (01)15414734014219(17)220630(10)7099144

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates