Reinforced Introducer Sets Maquet 7 Fr., Part Number:… recallZ-0932-2020 · Datascope Corporation · initiated Nov 20, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Datascope Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for compromised sterility in breached pouches of specific lots of Reinforced Introducer Sets Maquet 7 Fr., 7.5 Fr. and 8 Fr. IABs.
FDA root cause category: Package design/selection
What the company is doing
On December 11, 2019 the firm distributed Urgent Medical Device Recall Communication letters by FedEx 2 day delivery with signature proof of delivery stating: URGENT MEDICAL DEVICE RECALL REMOVAL MAQUET 7Fr., 7.5Fr. and 8Fr. REINFORCED INTRODUCER SETS PLEASE FORWARD THIS INFORMATION TO ALL CURRENT AND POTENTIAL USERS OF REINFORCED INTRODUCER SETS FOR MAQUET 7 Fr., 7.5 Fr. AND 8 Fr. INTRA-AORTIC BALLOON (IAB) WITHIN YOUR HOSPITAL OR FACILITY. IF YOU HAVE FURTHER DISTRIBUTED ANY OF THE AFFECTED PRODUCTS, FORWARD THIS INFORMATION TO THE RECIPIENT. Dear Risk Manager, Datascope / Maquet Getinge is initiating a voluntary Medical Device Recall Removal of the Reinforced Introducer Sets for Maquet 7 Fr., 7.5 Fr.and 8 Fr. IAB (Intra Aortic Balloon) due to a potential breach in the Mylar side (clear plastic) of the pouch, which could compromise device sterility. Identification of the issue: During the execution of a remediation protocol for re-verification of packaging, Datascope / Maquet Getinge found a small slit on the Mylar side (clear plastic) of the pouch of one device. Imprints of the sharp corners of the polystyrene strip used in packaging were observed around the slit, which likely caused the breach in the pouch compromising the sterility of the device. Risk to Health Adequate packaging and packaging materials are essential to help preserve the sterility of medical devices. Improperly sterilized medical devices may pose a serious risk of patient infection, including pyrogenic reaction, inflammatory response, and the possibility of a deadly infection. The probability of serious injury or illness occurring as a result of this potential issue is unlikely, however, likelihood is greater for immunocompromised patients. To date, Datascope / Maquet Getinge has not received any complaints, nor have any adverse events been reported resulting in serious illness or injuries caused by the Reinforced Introducer Set for Maquet 7 Fr., 7.5 Fr.
Recall details
- Recall number
- Z-0932-2020
- Status
- Terminated
- Initiated by the company
- Nov 20, 2019
- Recalling firm
- Datascope Corporation
- Quantity
- 102
- Product code
- DSP
- Distribution
- US: AK AL AZ CA CO CT FL GA IA ID IL IN KS KY LA MD MI MN MO MS NC ND NE NJ NY OH OK PA SC SD TN TX VA WA WI WV OUS: Switzerland, Netherlands, South Africa, Canada, Germany, Austria, Spain, Italy, Poland, Bharain, Malaysia
Product description
Reinforced Introducer Sets Maquet 7 Fr., Part Number: 0684 00 0403 06. It is an accessory to be used for percutaneous insertion of MAQUET Intra Aortic Balloon Catheters.
Code information (models, lots, serial numbers)
Part Number: 0684 00 0403 06 UDI: 10607567106656 (0684 00 0403 06) Lot Codes: 3000051658, TLT
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Reinforced Introducer Set for Linear 7.5Fr & Mega 7.5Fr 0684-00-0403-05Datascope Corp. · Introducer, Catheter2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 84521).
- Z-0933-2020 · TerminatedReinforced Introducer Sets Maquet 7.5 Fr., Part Number: 068400040305. It is an accessory to be used for percutaneous insertion of MAQUET IntraAortic Balloon Catheters.
- Z-0934-2020 · TerminatedReinforced Introducer Sets Maquet 8 Fr., Part Number: 068400040310. It is an accessory to be used for percutaneous insertion of MAQUET IntraAortic Balloon Catheters.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page