Skip to content

GETINGE 46-SERIES recallZ-1390-2020 · Maquet Cardiovascular Us Sales, Llc · initiated Jan 16, 2020

Class III Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Maquet Cardiovascular Us Sales, Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to a missing water intake for Deionized water on the Getinge 46-Series Washer Disinfector. There were no valve, hose, or clamps for this option included with the device therefore the units could not be properly installed. There have been no adverse events reported resulting in serious illness or injuries.

FDA root cause category: Employee error

What the company is doing

Getinge issued an Urgent Medical Device Recall Correction letter to U.S. consignees via FedEx Priority Overnight delivery with signature proof of delivery on January 16, 2020. Letter provides identification of the issue, health risk and action to take. Customers were advised to examine your inventory immediately, complete and sign the MEDICAL DEVICE RECALL - RESPONSE FORM to acknowledge that you have received this notification, return the completed form to Getinge Disinfection AB by e-mailing a scanned copy to [email protected] or by faxing the form to 1-800-574-5051. Should you have one or more Getinge 46-Series Washer Disinfectors (models 46-4-502 and 46-4-503) you will be contacted by a Getinge Service Representative to schedule on-site service of your device, free of charge, by a Getinge Service Representative. You can also arrange a visit by a Getinge trained or authorized service technician by contacting Getinge Customer Technical Support at (888) 9438872 (select option 1, then option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone). If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action.

Recall details

Recall number
Z-1390-2020
Recall class
Class III
Status
Terminated
Initiated by the company
Jan 16, 2020
Recalling firm
Maquet Cardiovascular Us Sales, Llc
Quantity
4 units
Product code
MEC
Distribution
U.S. Distribution: CA, CO, IA, MN, NY, SC, VA

Product description

Getinge 46-Series Washer Disinfector Model # 46-4, Part Number:46-4-502 Product Usage: Used in facilities such as Central Sterile Supply Departments, next to the Operating Room, in dental clinics or hospital wards to clean, thermally or chemically disinfect and dry surgical goods.

Code information (models, lots, serial numbers)

Serial Numbers: WAA090630, WAA090724, WAA091212, WAA091456, WAA091468, WAA092596. UDI Code 07340153700031

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 84885).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page