Dexcom G6 recallZ-1562-2020 · Dexcom Inc · initiated Feb 3, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Dexcom Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The company identified potential interference from hydroxyurea. Patient use of the anti-neoplastic drug may falsely elevate glucose readings on the Continuous Glucose Monitoring Systems.
FDA root cause category: Under Investigation by firm
What the company is doing
Starting in February 2020, the firm will add an "URGENT MEDICAL DEVICE CORRECTION" Notification to sensor packaging informing valued customers that the firm recently became aware that hydroxyurea, an anti-neoplastic drug used primarily in chemotherapy, falsely elevates glucose readings on its Continuous Glucose Monitoring (CGM) Systems. This may result in failure to treat hypoglycemia from missed low sensor glucose values and alerts or by causing other errors in diabetes management decisions (for example, hypoglycemia resulting from an insulin dose based on a falsely elevated sensor glucose reading). Some examples of the disease states that hydroxyurea is more commonly used in include: -Chronic Myelogenous Leukemia (CML) -Head and neck malignancies, and -Sickle cell disease The Recalling Firm is instructing its customers to not use their CGM System for diabetes treatment decisions if they are taking hydroxyurea. Customers are instructed to talk to their physician about alternative glucose monitoring approaches, and if they are unsure if they have been prescribed hydroxyurea. The Recalling Firm is currently updating their product labelling as following: "Hydroxyurea is a medication used in the treatment of diseases including cancer and sickle cell anemia; it is known to interfere with readings from your sensor. If you are taking hydroxyurea, your sensor glucose readings will be higher than your actual glucose, which could result in missed hypoglycemia alerts, or errors in diabetes management, such as giving yourself a higher dose of insulin due to falsely high sensor glucose values. The level of inaccuracy depends on the amount of hydroxyurea in your body. Do not use your Dexcom CGM System for diabetes treatment decisions if you are taking hydroxyurea. Talk to your physician about alternative glucose monitoring approaches. . . .Should you have any questions regarding hydroxyurea or alternative glucose monitoring approaches, please contact your healthcare
Recall details
- Recall number
- Z-1562-2020
- Status
- Terminated
- Initiated by the company
- Feb 3, 2020
- Recalling firm
- Dexcom Inc
- Quantity
- 275583
- Product code
- QBJ
- FDA submissions
- K191450
- Distribution
- World-wide Distribution
Product description
Dexcom G6 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracking patterns in persons (age 18 and older) with diabetes.
Code information (models, lots, serial numbers)
Device Listing Number: D319131 Model: STS-OR-003 Description: G6 Sensor Kit 3 Pack GUDID Registered GTIN: 50386270000250 Lot Number: All Model: STS-OE-003 Description: G6 Sensor Kit 3 Pack Retail GUDID Registered GTIN: 00386270000866 Lot Number: All Model: STS-OM-003 Description: G6 Sensor Kit 3 Pack Medicare GUDID Registered GTIN: 00386270001047 Lot Number: All Model: STS-OM-003 Description: G6 Sensor Kit 3 Pack Pro Q GUDID Registered GTIN: 70386270000254 Lot Number: All
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Dexcom G6 Sensor Applicator Assembly 9500-45Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall4 deaths114,678 reports
- Dexcom Discovery Sensor 3 Pack STS-DE-003Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 84891).
- Z-1560-2020 · TerminatedDexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends a…
- Z-1561-2020 · TerminatedDexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracki…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.