Skip to content

ENCOR recallZ-1431-2020 · Bard Peripheral Vascular Inc · initiated Jan 3, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Bard Peripheral Vascular Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Lot numbers and products inadvertently not included in the scope of the previous recall. Original recall was conducted due to an increase in probe failures for leaks, suction issues, and failure to obtain samples.

FDA root cause category: Component change control

What the company is doing

The recalling firm issued letters dated 1/3/2020 to each of the U.S. consignees via FedEx with proof of delivery notification. The letter identified the affected product, problem and actions to be taken. For questions call 1-800-321-4254 Option #5.

Recall details

Recall number
Z-1431-2020
Status
Terminated
Initiated by the company
Jan 3, 2020
Recalling firm
Bard Peripheral Vascular Inc
Quantity
12,875 units
Product code
KNW
FDA submissions
K040842, K051158
Distribution
Distribution was made to AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OH, OK, OR, PA, SC, TX, VA, and WV. There was government distribution and no military distribution. Foreign distribution was made to Canada, Australia, Austria, Belgium, China, Czech Republic, Germany, France, Georgia, Greece, Hong Kong, Israel, Italy, Japan, Korea, Netherlands, New Zealand, Poland, Portugal, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, and United Kingdom.

Product description

Bard Biopsy EnCor Probe, Stereotactic/Ultrasound 7G, REF number ECP017G, packaged individually in sterile pouches, 5 pouches/carton, Single Use, Rx, Sterile.

Code information (models, lots, serial numbers)

Lot numbers VTCT0268, VTCT0269, VTCT0270, VTCT0271, VTCU0340, and VTCU0341.   UDI numbers: (01)00801741086281(17)200601(10)VTCT0268, (01)00801741086281(17)200601(10)VTCT0269, (01)00801741086281(17)200601(10)VTCT0270, (01)00801741086281(17)200601(10)VTCT0271, (01)00801741086281(17)200601(10)VTCU0340, and (01)00801741086281(17)200601(10)VTCU0341.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

4 devices cleared under K040842, K051158 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 84929).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page