ADVIA Chemistry Total Bilirubin_2 assay recallZ-1539-2020 · Siemens Healthcare Diagnostics, Inc. · initiated Feb 13, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy
FDA root cause category: Under Investigation by firm
What the company is doing
Siemens Healthineers issued Urgent Medical Device Correction (UMDC # ACHC-20-02.A.US, ACHC-20-02.A.US.AC, ACHC-20-02.A.US.DM, ACHC-20-02.A.US.DV) and Urgent Field Safety Notice (UFSN # ACHC-20-02.A.OUS, ACHC-20-02.A.OUS.AC, ACHC-20-02.A.OUS.DM, ACHC-20-02.A.OUS.DV) on February 13, 2020 to all Atellica CH, ADVIA CH, Dimension, and Dimension Vista customers to inform them of the issue, the potential risk to health, and actions to be taken by the customer. Customers were instructed to:Review the information on eltrombopag interference (ADVIA CH and Atellica CH bilirubins) or use of total bilirubin assay is not recommended for patients on eltrombopag therapy (Dimension and Dimension Vista). Review the letter with their Medical Director Complete and return the Field Correction Effectiveness Check Form attached to the letter within 30 days. If they have received any complaints of illness or adverse events associated with the product, immediately contact their local Siemens Customer Care Center or their local Siemens Technical Support Representative.
Recall details
- Recall number
- Z-1539-2020
- Status
- Terminated
- Initiated by the company
- Feb 13, 2020
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 5666 US and 52202 OUS
- Product code
- JFM
- FDA submissions
- K170065
- Distribution
- Nationwide Foreign: Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, United Kingdom, Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Azerbaijan, Bahamas, Bahrain, Bangladesh, Benin, Bosnia Herzegovina, Botswana, Brazil, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Congo, Ecuador, Egypt, French Guinea, Georgia, Guadeloupe, Hong Kong, India, Indonesia, Iraq, Iran, Israel, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Lesotho, Libya, Macedonia, Malawi, Malaysia, Mali, Mexico, Montenegro, Morocco, Mozambique, Myanmar, New Zealand, Nigeria, Oman, Pakistan, Peru, Philippines, Qatar, Republic if Yemen, Republic Korea, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, South Africa, Swaziland, Taiwan, Tanzania, Thailand, Tunisia, Turkmenistan, United Arab Emirates, Ukraine, Uruguay, Venezuela, Vietnam, Zimbabwe
Product description
ADVIA Chemistry Total Bilirubin_2 assay- SMN: 10341115 (40 mL) & 10341113 (70 mL). Product Usage: ADVIA TBIL_2 is for in vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma of adults and neonates on the ADVIA Chemistry systems. Measurement of total bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gallbladder block. A total bilirubin measurement in newborn infants is intended to aid in indicating the risk of bilirubin encephalopathy (kernicterus).
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
ADVIA CH TBIL_2 (40 mL) 492094 (01)00630414513072(10)492094(17)20210828 ADVIA CH TBIL_2 (40 mL) 486525 (01)00630414513072(10)486525(17)20210728 ADVIA CH TBIL_2 (40 mL) 478149 (01)00630414513072(10)478149(17)20210428 ADVIA CH TBIL_2 (40 mL) 472677 (01)00630414513072(10)472677(17)20210328 ADVIA CH TBIL_2 (40 mL) 469083 (01)00630414513072(10)469083(17)20210128 ADVIA CH TBIL_2 (40 mL) 462464 (01)00630414513072(10)462464(17)20201128 ADVIA CH TBIL_2 (40 mL) 455486 (01)00630414513072(10)455486(17)20200828 ADVIA CH TBIL_2 (40 mL) 452851 (01)00630414513072(10)452851(17)20200728 ADVIA CH TBIL_2 (40 mL) 447317 (01)00630414513072(10)447317(17)20200628 ADVIA CH TBIL_2 (40 mL) 436498 (01)00630414513072(10)436498(17)20200328 ADVIA CH TBIL_2 (70 mL) 492102 (01)00630414513058(10)492102(17)20210828 ADVIA CH TBIL_2 (70 mL) 492111 (01)00630414513058(10)492111(17)20210828 ADVIA CH TBIL_2 (70 mL) 486533 (01)00630414513058(10)486533(17)20210728 ADVIA CH TBIL_2 (70 mL) 478154 (01)00630414513058(10)478154(17)20210428 ADVIA CH TBIL_2 (70 mL) 478159 (01)00630414513058(10)478159(17)20210428 ADVIA CH TBIL_2 (70 mL) 493710 (01)00630414513058(10)493710(17)20210428 ADVIA CH TBIL_2 (70 mL) 490263 (01)00630414513058(10)490263(17)20210328 ADVIA CH TBIL_2 (70 mL) 490264 (01)00630414513058(10)490264(17)20210328 ADVIA CH TBIL_2 (70 mL) 469088 (01)00630414513058(10)469088(17)20210128 ADVIA CH TBIL_2 (70 mL) 469094 (01)00630414513058(10)469094(17)20210128 ADVIA CH TBIL_2 (70 mL) 462470 (01)00630414513058(10)462470(17)20201128 ADVIA CH TBIL_2 (70 mL) 462477 (01)00630414513058(10)462477(17)20201128 ADVIA CH TBIL_2 (70 mL) 455496 (01)00630414513058(10)455496(17)20200828 ADVIA CH TBIL_2 (70 mL) 455501 (01)00630414513058(10)455501(17)20200828 ADVIA CH TBIL_2 (70 mL) 452858 (01)00630414513058(10)452858(17)20200728 ADVIA CH TBIL_2 (70 mL) 452863 (01)00630414513058(10)452863(17)20200728 ADVIA CH TBIL_2 (70 mL) 452870 (01)00630414513058(10)452870(17)20200728 ADVIA CH TBIL_2 (70 mL) 447325 (01)00630414513058(10)447325(17)20200628 ADVIA CH TBIL_2 (70 mL) 447332 (01)00630414513058(10)447332(17)20200628 ADVIA CH TBIL_2 (70 mL) 436510 (01)00630414513058(10)436510(17)20200328 ADVIA CH TBIL_2 (70 mL) 436515 (01)00630414513058(10)436515(17)20200328
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- ADVIA® Chemistry Total Bilirubin_2 TBIL_2 Reagents 10341113Siemens Healthcare Diagnostics Inc. · BILIRUBIN (TOTAL AND UNBOUND) IN THE NEONATE TEST SYSTEM1 recall0 reports
- ADVIA® Chemistry TBIL_2 Total Bilirubin_2 Reagents 10341115Siemens Healthcare Diagnostics Inc. · ENZYMATIC METHOD, BILIRUBIN1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 85037).
- Z-1536-2020 · TerminatedAtellica CH Direct Bilirubin_2 assay In vitro Diagnostic for the determination of direct bilirubin SMN: 11097532 - Product Usage: ADVIA DBIL_ 2 is for in vitro diagnostic use in th…
- Z-1537-2020 · TerminatedAtellica CH Total Bilirubin_2 assay-In vitro Diagnostic for the determination of total bilirubin SMN: 11097531 - Product Usage: Atellica TBil_2 assay is for in vitro diagnostic us…
- Z-1538-2020 · TerminatedADVIA Chemistry Direct Bilirubin_2 assay-SMN: 10316610 (20 mL) & 10341114 (70 mL) - Product Usage: ADVIA DBIL_ 2 is for in vitro diagnostic use in the quantitative determination o…
- Z-1540-2020 · TerminatedDimension Total Bilirubin Flex reagent cartridge -In vitro Diagnostic for the determination of total bilirubin. SMN: 10444957 Product Usage: Dimension TBI method for the clinic…
- Z-1541-2020 · TerminatedDimension Vista Total Bilirubin Flex reagent cartridge -In vitro Diagnostic for the determination of total bilirubin. SMN: 10445146 Product Usage: Dimension Vista TBIL method is…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.