RayStation/RayPlan 8B Service Pack 2, Build Number 8.1.2.5… recallZ-1466-2020 · Raysearch Laboratories AB · initiated Feb 20, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Raysearch Laboratories AB or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.
FDA root cause category: Under Investigation by firm
What the company is doing
An Urgent Field Safety Notice, Medical Device Correction notification letter dated 2/19/20 was emailed to the customer. ACTIONS TO BE TAKEN BY THE USER " Either do not use the RayTreat tabletop propagation functionality or make sure to always perform imaging and registration correction before each treatment session. " When using RayTreat in combination with the ProNova machine, do not use the acquired imaging positions in the preparation tab as the truth about where an image was actually taken. Please educate all users including the planning, Quality Assurance and plan delivery staff about this workaround. Inspect your product and identify all installed units with the above software version number(s), then confirm you have read and understood this notice by replying to the notification email. SOLUTION This issue will be resolved in the next version of RayStation, scheduled for market release in March 2020 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use. TRANSMISSION OF THIS NOTICE This notice needs to be passed on to all those who need to be aware within your organization. Please maintain awareness of this notice as long as any version of RayStation affected by this issue is in use to ensure effectiveness of the workaround.
Recall details
- Recall number
- Z-1466-2020
- Status
- Terminated
- Initiated by the company
- Feb 20, 2020
- Recalling firm
- Raysearch Laboratories AB
- Quantity
- 1 system in the U.S.
- Product code
- MUJ
- FDA submissions
- K190387
- Distribution
- The products were distributed to the following US states: TN
Product description
RayStation/RayPlan 8B Service Pack 2, Build Number 8.1.2.5, stand-alone software treatment planning system
Code information (models, lots, serial numbers)
UDI 07350002010235
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- RayStation 8.1.2Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment8 recalls0 reports
Other devices under the same FDA submission
2 devices cleared under K190387 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- RayStation 9Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment12 recalls0 reports
- RayStation 8.1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment8 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 85065).
- Z-1462-2020 · TerminatedRayStation/RayPlan 8A, Build Number 8.0.0.61, stand-alone software treatment planning system
- Z-1463-2020 · TerminatedRayStation/RayPlan 8A Service Pack 1, Build Number 8.0.1.10, stand-alone software treatment planning system
- Z-1464-2020 · TerminatedRayStation/RayPlan 8B, Build Number 8.1.0.47, stand-alone software treatment planning system
- Z-1465-2020 · TerminatedRayStation/RayPlan 8B Service Pack 1, Build Number 8.1.1.8, stand-alone software treatment planning system
- Z-1467-2020 · TerminatedRayStation/RayPlan 9A, Build Number 9.0.0.113, stand-alone software treatment planning system
- Z-1468-2020 · TerminatedRayStation/RayPlan 9B, Build Number 9.1.0.933, stand-alone software treatment planning system
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page