Skip to content

Guardian Connect App CSS7200 iOS and Guardian Connect… recallZ-2106-2020 · Medtronic Inc. · initiated May 20, 2019

Class II Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Customers using the firm's continuous glucose monitoring system application on an iPhone, iPad or iPod Touch with ios software version 12, 12.1, or 12,2 are likely to experience a shortened transmitter battery life (approximately 4-5 days instead of the normal 7 days or more) after a full charge.

FDA root cause category: Software design

What the company is doing

Urgent Medical Device Safety Notification Letters were sent to patients, health care providers (HCPs), distributors and clinical via email beginning on May 20, 2019 and via USPS certified mail beginning on May 22, 2019 notifying them of a potential issue with the transmitter battery life of the Connect Continuous Glucose Monitoring System. Customers using an Apple device with iOS version 12 or higher were advised to update to the latest version (3.2.4) of the Guardian Connect app on their iPhone as soon as possible. Customers not using iOS 12 or higher were encouraged to update to the latest version of the Guardian Connect app. Customers receiving emails were requested to submit email acknowledgements indicating they have read and understood the instructions provided in the Urgent Safety Notice Letter clicking on the button as evidence of receipt. Customers receiving paper-copy notifications were instructed to fill out a response card indicating they had read and understood the notice, and mail the card back with a pre-paid envelope. Customers with invalid email and/or physical addresses will be contacted by telephone. Customers with questions were encouraged to contact 24-Hour Technical Support Team at 1-800-646-4633.

Recall details

Recall number
Z-2106-2020
Recall class
Class II
Status
Open, Classified
Initiated by the company
May 20, 2019
Recalling firm
Medtronic Inc.
Quantity
23,978
Product code
MDS
FDA submissions
P160007
Distribution
US Nationwide distribution.

Product description

Guardian Connect App CSS7200 iOS and Guardian Connect Transmitter GST4C used on the iPhone, iPad and iPod Touch devices.

Code information (models, lots, serial numbers)

Model Number CSS7200 (Guardian Connect App: iOS); Catalog Numbers: MMT-7821 (Guardian Connect Transmitter - GST4C); UPN - 00763000186333.  There are no assigned lot or serial numbers.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under P160007 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page