Oncomine Dx Target Test RNA Control- IVD for detection of… recallZ-1705-2020 · Life Technologies Corporation · initiated Mar 16, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Life Technologies Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Extended Expiration Date prior to clearance
FDA root cause category: Incorrect or no expiration date
What the company is doing
Thermo Fisher issued notification on 3/16/20 via email followed by hard copy on 3/18/20, stating reason for recall, health risk and action to take: Review the expiration date of the products referenced in the table included in the Customer Notification and take the following actions: "Notify all Oncomine Dx Target Test users in your facility of this issue. "Review the expiration date of the products referenced in the table above and using Attachment 1 of this letter, take the following actions. Attachment 1 provides the corrected expiration dates per lot. For products that are past the 9-month corrected expiration date per Attachment 1: Discontinue use of the product, record the kit lot numbers and quantities impacted on the attached Customer Response Sheet, and email the completed response sheet to [email protected]. For products that are still within the 9-month corrected expiration date per Attachment 1: Correct the expiration dating on the product labelling within your inventory by reviewing Attachment 1 to determine the correct expiration date per kit lot. Strike through the incorrect expiration date and add the new one using permanent ink or amend the label by adding an additional label correcting the expiration date on the labelling. The updated labelled product can continue to be used until the 9-month expiration date, after which any remaining product should be discarded. Note: If affected product was used past the 9-month corrected expiration date per Attachment 1, but before the 12-month expiration date on the kit labels, there is no anticipated risk of generating inaccurate/incorrect results because the system will inform the user if the assay did not meet run passing criteria and test results wont be generated. Therefore, no action on your part is required. If you have any technical questions or concerns, please contact Technical Support at [NA --800-955-6288 option 4 or [email protected]]
Recall details
- Recall number
- Z-1705-2020
- Status
- Terminated
- Initiated by the company
- Mar 16, 2020
- Recalling firm
- Life Technologies Corporation
- Quantity
- 66 units
- Product code
- PQP
- FDA submissions
- P160045
- Distribution
- US Nationwide distribution.
Product description
Oncomine Dx Target Test RNA Control- IVD for detection of DNA and RNA isolated from non-small cell lung cancer (NSCLC) tumor specimens, Catalog Number: A32443
Code information (models, lots, serial numbers)
Lot Codes: S119_2174739 S119_2108428 P119_2149305 N119_2078114 S218_2069395 V118_1991154 T118_1990946 S118_1988603 P118_1996940 S117_1988306 UDI: (01)10190302006064
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Oncomine™ Dx Target RNA Control Kit A32443Life Technologies Corporation · Next generation sequencing oncology panel, somatic or germline variant detection system1 recall0 reports
Other devices under the same FDA submission
5 devices cleared under P160045 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Oncomine™ Dx Target Test A32441Life Technologies Corporation · Next generation sequencing oncology panel, somatic or germline variant detection system1 recall1 report
- Oncomine™ Dx Target Test User Guides and Assay Definition File A32461Life Technologies Corporation · Next generation sequencing oncology panel, somatic or germline variant detection system1 recall0 reports
- Oncomine™ Dx Target DNA Control Kit A32442Life Technologies Corporation · Next generation sequencing oncology panel, somatic or germline variant detection system1 recall0 reports
- Ion Torrent™ Dx No Template Control Kit A32444Life Technologies Corporation · Next generation sequencing oncology panel, somatic or germline variant detection system1 recall0 reports
- Oncomine™ Dx Target Test A32441Life Technologies Corporation · Next generation sequencing oncology panel, somatic or germline variant detection system1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 85203).
- Z-1704-2020 · TerminatedOncomine Dx Target Test DNA Control-IVD for detection of DNA and RNA isolated from non-small cell lung cancer (NSCLC) tumor specimens, Catalog Number: A32442.
- Z-1706-2020 · TerminatedOncomine Dx Target Test RNA/DNA Panel- IVD for detection of DNA and RNA isolated from non-small cell lung cancer (NSCLC) tumor specimens, Catalog Number: A32441
- Z-1707-2020 · TerminatedIon Torrent Dx No Template Control Kit- IVD for detection of DNA and RNA isolated from non-small cell lung cancer (NSCLC) tumor specimens Catalog Number: A32444
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page