Skip to content

Stablecut recallZ-1676-2020 · Synvasive Technology Inc · initiated Jul 27, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Synvasive Technology Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Mislabeled: Actual blade width is 19 mm but the label indicates a width of 25 mm.

FDA root cause category: Labeling Change Control

What the company is doing

Zimmer Biomet issued Urgent Medical Device Recall letter dated 7/27/18 stating reason for recall, health risk and action to take: Review this notification and ensure that affected personnel are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your facility. a. Complete Attachment 1 Inventory Return Certification Form and send to [email protected] within three (3) days. b. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. c. Request an RGA number for each unit by calling 800-830-0970 or emailing [email protected]. d. Mark RECALL and RGA Number on the outside of the returned cartons. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 800-830-0970 between 8:00 am and 5:00pm EST, Monday through Friday

Recall details

Recall number
Z-1676-2020
Status
Terminated
Initiated by the company
Jul 27, 2018
Recalling firm
Synvasive Technology Inc
Quantity
266 units
Product code
GFA
Distribution
US distribution to states of AL and NY

Product description

ZIMMER Oscillating Saw Blade -Cutting 75mm x 19mm x 1.00/1.27mm.intended for bone or cartilage in orthopedic procedures. Manufactured By: Synvasive Technology, Inc., a Subsidiary of Zimmer

Code information (models, lots, serial numbers)

Lot Number/UDI: 265474 (01)00846795028320(17)220414(10)265474 304117 (01)00846795028320(17)230131(10)304117 317918 (01)00846795028320(17)230531(10)317918

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page