Skip to content

VACUETTE Tube 2ml 9NC Coagulation Sodium Citrate 3.2%. Item… recallZ-1714-2020 · Greiner Bio-One North America, Inc. · initiated Mar 12, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Greiner Bio-One North America, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Concerned lot of VACETTE 2mL 9NC Coagulation Tubes were shown to have inconsistent amounts of the 2.3% sodium citrate.

FDA root cause category: Process change control

What the company is doing

Greiner Bio-One North America Inc., notified customers on 03/12/2020 via "Urgent Product Recall" letter. The recall letter identified the affected products and requested the users to stop using the product and complete the product disposition site confirmation form. Consignees were asked to discard/destroy the affected products. If you have additional questions, please call our Technical Service at 800-815-8112.

Recall details

Recall number
Z-1714-2020
Status
Terminated
Initiated by the company
Mar 12, 2020
Recalling firm
Greiner Bio-One North America, Inc.
Quantity
1,000,800 units.
Product code
GIM
FDA submissions
K971221
Distribution
US nationwide distribution in the states of: NJ, IN, TN, MI, NC, WA, KS, CA, LA, IL, MD, NY, OH, TX, GA, KY, PA, CO, MA, PR, FL, AR, OK and NE.

Product description

VACUETTE Tube 2ml 9NC Coagulation Sodium Citrate 3.2%. Item no. 454322. Lot no. B1908355.

Code information (models, lots, serial numbers)

Item no. 454322. Lot no. B1908355. DI no. (01)29120017574488 (17) 200809 (10)B1908355.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

2 devices cleared under K971221 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.