Skip to content

ARCHITECT recallZ-1728-2020 · Abbott Gmbh & Co. KG · initiated Mar 9, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Gmbh & Co. KG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

An erroneous concentration value for hepatitis A virus was utilized during manufacture, leading to a decreased amount of antigen used for the impacted lots.

FDA root cause category: Mixed-up of materials/components

What the company is doing

Abbott GmbH notified customers on 03/09/2020 via "Product Recall Urgent Immediate Action Required" letter. The recall letter identified the affected products and requested the customers to Immediately discontinue use of, and destroy, any remaining inventory of these ARCHITECT HAVABG reagent lots.

Recall details

Recall number
Z-1728-2020
Status
Terminated
Initiated by the company
Mar 9, 2020
Recalling firm
Abbott Gmbh & Co. KG
Quantity
7,347 kits
Product code
LOL
FDA submissions
K113704
Distribution
US nationwide including the states of AL, AR, AZ, CA, CO, CT DE, FL, GA, ID, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NJ, NM NV, NY, OH, OK, OR, PA, RH, SC, TN, TZ, UT, VA, WA, WI, WV AND PR.

Product description

ARCHITECT HAVAB G Reagent Kit. List Number 6L27 25.

Code information (models, lots, serial numbers)

List Number 6L27 25. Lot No: 03429BE00, 06172BE00, 08073BE00, 10353BE00. UDI: (01)00380740011185 (17)200320(10)03429BE00, (01)00380740011185 (17)200604(10)06172BE00, (01)00380740011185 (17)200819(10)08073BE00, (01)00380740011185 (17)201023(10)10353BE00.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K113704 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page