VITROS Immunodiagnostic Products TT4 Reagent Pack, Product… recallZ-1834-2020 · Ortho Clinical Diagnostics · initiated Mar 10, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ortho Clinical Diagnostics or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The specified reagent packs exhibit an increase in the frequency of unsuccessful calibrations.

FDA root cause category: Under Investigation by firm

What the company is doing

IMPORTANT PRODUCT CORRECTION NOTIFICATION letters dated 3/10/20 were sent to customers. REQUIRED ACTIONS " If you have VITROS TT4 Reagent Packs Lot 2340 in your inventory, Ortho will replace or credit your account as indicated on your Confirmation of Receipt form. NOTE: As a reminder, if you have a successful calibration, and quality control results are acceptable, it is acceptable to continue using the affected lot. " Post this notification by each system that processes VITROS TT4 Reagent Packs, Lot 2340. " Complete the Confirmation of Receipt form now to register receipt of the notification. All customers should return the completed form by March 25, 2020. " If you are satisfied with the product at this time but want Lot 2340 replaced in the future, you may resubmit the Confirmation of Receipt form at that time, indicating the quantity that requires replacement. " In the event of future occurrences, retain a copy of the Confirmation of Receipt form and re-submit if needed. " Please forward this notification if the product was distributed outside of your facility. Contact Information We apologize for the inconvenience this will cause your laboratory. If you have further questions, please contact Ortho Care" Technical Solutions Center at 1-800-421-3311.

Recall details

Recall number
Z-1834-2020
Status
Terminated
Initiated by the company
Mar 10, 2020
Recalling firm
Ortho Clinical Diagnostics
Quantity
3181
Product code
KLI
FDA submissions
K160495
Distribution
US Nationwide. Australia, Bermuda, Belgium, Brazil, Canada, Chile, China, Columbia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, UK.

Product description

VITROS Immunodiagnostic Products TT4 Reagent Pack, Product Code 8744468

Code information (models, lots, serial numbers)

UDI: 10758750005147; Lot Number: 2340