EndoWrist SP recallZ-2131-2020 · Intuitive Surgical, Inc. · initiated Aug 19, 2019

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Intuitive Surgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Some cameras were found to have residual soil adhered to the tip of the camera after reprocessing which may affect image quality resulting in needing to switch to a second camera or converting to another surgical modality. If residual soil makes contact with the patient, exposure could result in sensitization making future component blood therapy or tissue typing more challenging; additionally if a small piece of inert epoxy were to be retained, scar tissue could form.

FDA root cause category: Process control

What the company is doing

Urgent Medical Device Recall Letters (Field Safety Notices) were hand-delivered to U.S. customers on August 19, 2019 and OUS customers on August 20, 2019. Customers were asked to locate affected camera(s) in their inventory and inspect per the instructions attached to the Field Safety Notice throughout the life of the camera(s). If residual soil is present, customers were to return the affected camera(s) to the manufacturer using the standard Return Materials Authorization (RMA) process. Additionally, customers were requested to: iInform necessary hospital personnel about this issue; complete the attached Acknowledgement Form and return it via fax to Intuitive as instructed on the form; and to retain a copy of this letter and the Acknowledgement Form for their files. Customers were informed that an Intuitive Representative would schedule a site visit to review the attached inspection instructions, and that the firm will provide a replacement camera, at no charge, for returned affected product. Customers requiring further information or support may contact their Clinical Sales Representative or contact Intuitive Customer Service at the following numbers: US Customer Service (800) 876-1310, Option 3 (7 AM to 5 PM EST) or [email protected], or South Korea Customer Service: + 82-02-3271-3200, Option 3 ( 9AM to 6 PM KSTJ)

Recall details

Recall number
Z-2131-2020
Status
Open, Classified
Initiated by the company
Aug 19, 2019
Recalling firm
Intuitive Surgical, Inc.
Quantity
305 units
Product code
NAY
FDA submissions
K173906
Distribution
Distribution US nationwide and South Korea.

Product description

da Vinci SP Camera; PN 430060-33; Model Number: 430060-33. For endoscopic surgery.

Code information (models, lots, serial numbers)

Model Number: 430060-33; UDI Number: 00886874114384; All serial numbers.