Skip to content

Xhibit recallZ-2261-2020 · Spacelabs Healthcare, Inc. · initiated Aug 30, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Spacelabs Healthcare, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm received reports of loss of audio alarm after a power failure or cable disconnection. The loss of audio for a telemetry patient could delay the recognition of an alarm condition.

FDA root cause category: Software Manufacturing/Software Deployment

What the company is doing

Urgent Medical Device Correction letters were sent to U.S. customers on August 30, 2019 via priority service, and to foreign customers via email on September 6, 2019. The letter informed customers of the conditions under which alarm audio may be lost, and actions the customer /user can take to minimize risk to patients or users. Customers were requested to weigh the benefits versus the risks when deciding whether or not to continue to use Xhibit Central with software versions 1.3.1, 1.3.2, 1.3.3, and 1.3.4. Customers experiencing a loss of audio were informed that a complete reboot of the Xhibit Central will return the device to normal operation. Customers were requested to circulate this notification to all persons affected by the information it contains. Customers were also informed that the manufacturer would contact them to schedule a time for Spacelabs to update their Xhibit Central software. Customers with questions, requiring information, or technical assistance may contact the Technical Support group at Spacelabs Healthcare, Inc. at 1-800-522-7025 and select 2 for Technical Support.

Recall details

Recall number
Z-2261-2020
Status
Terminated
Initiated by the company
Aug 30, 2019
Recalling firm
Spacelabs Healthcare, Inc.
Quantity
2197 units
Product code
MHX
FDA submissions
K122146
Distribution
Worldwide distribution - US Nationwide distribution including in the 40 states and Puerto Rico, and multiple countries including Canada.

Product description

Xhibit Central, Model No. 96102 - Product Usage: intended use is to provide clinicians with central monitoring of adult, pediatric and neonatal patient data of patients connected to networked Spacelabs Healthcare patient monitors and telemetry transmitters.

Code information (models, lots, serial numbers)

Model No. 96102 Xhibit Central; Software Versions: 1.3.1, 1.3.2, 1.3.3 and 1.3.4; UDI: 10841522100345

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

2 devices cleared under K122146 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page