Custom Sheath Introducer Kit REF K21-00031 - Product Usage:… recallZ-1956-2020 · Merit Medical Systems, Inc. · initiated Sep 12, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Merit Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
labeling error: Due to a manufacturing issue, product package was incorrectly labeled with the incorrect sheath size and the incorrect sheath assembly was included in the product packaging. The use of the incorrect kit components may result in vasoconstriction, moderate hemorrhage, and/or embolism.
FDA root cause category: Process control
What the company is doing
A verbal (telephone) notification attempt was made to the customer on September 12th and 13th. These attempts resulted in a voicemail informing the customer of a voluntary product recall and requesting a call back. A hospital employee was reached by telephone on September 17, 2019 and notified of the product recall. The notification included product identification information, instruction to immediately quarantine any devices and to discontinue use and to return product to Merit per the instructions contained in the Customer Response Form (CRF). The CRF and details of the recall were emailed to the hospital per customer request.. Merit Medical Systems, Inc. is conducting a recall of the K21-00031 Custom Prelude IDeal Kit, lot number H1583914, due to the kit including a Prelude EASE sheath assembly rather than the labeled Prelude IDeal sheath assembly. Additionally, the kit has a labeling error in which the unit and carton labeling indicate a 6 French color (green) rather than the correct 4 French color (red). Merit Medical has not received any complaints from customers or any reports of patient harm or injury as a result of this issue. Return the affected products to Merit via Federal Express 2nd Day Air (Acct# 112952756), include the assigned RMA number. Send to Field Assurance Department Merit Medical Systems, Inc. 1600 West Merit Parkway South Jordan, UT 84095 Attn: RMA# (found in contact information table) For further information contact Merit Customer service at (801) 208-4381 or by email at [email protected].
Recall details
- Recall number
- Z-1956-2020
- Status
- Terminated
- Initiated by the company
- Sep 12, 2019
- Recalling firm
- Merit Medical Systems, Inc.
- Quantity
- 15 kits
- Product code
- DYB
- FDA submissions
- K173750
- Distribution
- US Nationwide distribution in the state of MI.
Product description
Custom Sheath Introducer Kit REF K21-00031 - Product Usage: intended use is to provide access and facilitate the percutaneous introduction of various devices into vein.
Code information (models, lots, serial numbers)
Catalog Number: K21-00031, Lot Number: H1583914, UDI: 00884450426357
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Prelude IDeal™ Hydrophilic Sheath 00884450426357Merit Medical Systems, Inc. · INTRODUCER, CATHETER1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page