Synaptive ImageDrive Clinical recallZ-1759-2020 · Synaptive Medical Inc · initiated Feb 26, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Synaptive Medical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A software defect negatively affects the quality of images taken during diagnostic imaging (mammograms) which may result in misdiagnosis.

FDA root cause category: Under Investigation by firm

What the company is doing

The firm initiated this field corrective action by letter on 02/26/2020. The consignees are not required to discontinue using the device nor return the device to Synaptive Medical, a workaround is to be followed as indicated in the customer notification letter. The actions to be taken by consignees with the affected product are as follows: 1. Acknowledge receipt of the notification using the acknowledgment and receipt form that the firm has provided. 2. a) If ClearCanvas Workstation is NOT used for viewing images on removeable media such as CDs, and this issue is encountered, configure the PACS or DICOM device that is sending images to ClearCanvas Workstation to send uncompressed images, or images compressed with a compression algorithm other than JPEG 2000. b.) If ClearCanvas Workstation IS used for viewing images on removeable media such as CDs OR if it is not possible to change the sending device to send uncompressed images, contact Synaptives Product Support Recall Line at 1.866.853.8001 to request assistance to correct the device. Note that if the software is subsequently re-installed, the device will need to be corrected again. Contact Synaptive to repeat the correction.

Recall details

Recall number
Z-1759-2020
Status
Terminated
Initiated by the company
Feb 26, 2020
Recalling firm
Synaptive Medical Inc
Quantity
28 units
Product code
LLZ
FDA submissions
K153284
Distribution
US

Product description

Synaptive ImageDrive Clinical

Code information (models, lots, serial numbers)

ImageDrive Clinical with part number SYN-0578- all versions of the software. With unique device identifier (UDI) #(01) 00670082000139 (10)13x1

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K153284 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 85216).