PowerPort ClearVUE recallZ-1964-2020 · Bard Peripheral Vascular Inc · initiated Jan 15, 2020

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Bard Peripheral Vascular Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The catheter locks for the lot may contain an 8Fr catheter lock instead of a 6Fr catheter lock per device labeling.

FDA root cause category: Employee error

What the company is doing

The recalling firm issued letters dated 1/15/2020 via FedEx with proof of delivery notification. The devices were to be returned.

Recall details

Recall number
Z-1964-2020
Status
Open, Classified
Initiated by the company
Jan 15, 2020
Recalling firm
Bard Peripheral Vascular Inc
Quantity
200 units
Product code
LJT
FDA submissions
K122899
Distribution
US Nationwide distribution including in the states of AZ, CA, MA, MD, NJ, NY, and TX. There was no foreign/government/military distribution.

Product description

BARD Access Systems PowerPort ClearVUE Slim Implantable Port With Smooth Septum and Attachable 6 F Polyurethane Open-Ended Single-Lumen Venous Catheter, Product #5676300 - Product Usage: All materials are biocompatible, can be used with virtually all injectable solutions intended for medicinal use, including the power injection of contrast media.

Code information (models, lots, serial numbers)

Lot #RECX3530, Exp. date 4/30/2020, UDI (01)00801741110771(17)200430(10)RECX3530.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates