Extremities, Large Kit Assy, US, 15/3 - Product Usage: is… recallZ-2389-2020 · Medtronic Sofamor Danek USA Inc · initiated Apr 24, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Sofamor Danek USA Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Product may not maintain its setting characteristics for the labeled shelf-life duration. Use may lead to longer time for product to set at the time of implantation which may increase the required surgical and anesthesia time for the patient. If the device seeps or creeps from its intended position prior to setting and creates incomplete void or gap filling, possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
FDA root cause category: Under Investigation by firm
What the company is doing
Medtronic notified customers on about 04/24/2020 via "URGENT: Voluntary Market Removal (Recall)" letter. Customers were informed that Medtronic was notified by Skeletal Kinetics LLC that product that may not maintain its setting characteristics for the labeled shelf-life duration of the product, which may lead to a longer period of time required for the product to set at the time of implantation. Instructions included to not use the product and locate and removed the impacted product from inventory, complete the provided Customer Confirmation Form and return it via email to [email protected], and contact your Medtronic Sales Representative to coordinate the return and replacement of affected product.
Recall details
- Recall number
- Z-2389-2020
- Status
- Terminated
- Initiated by the company
- Apr 24, 2020
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Quantity
- 41 units
- Product code
- HRX
- FDA submissions
- K010246, K041454, K981251
- Distribution
- Nationwide distribution to CA, FL, IL, IN, KY, MN, MO, MT, OH, PA, TX, VA, WA.
Product description
Extremities, Large Kit Assy, US, 15/3 - Product Usage: is an injectable, moldable and biocompatible calcium phosphate bone void filler. The single-use Medtronic Bone Void Filler Kit contains the necessary components for mixing of the bone void filler.
Code information (models, lots, serial numbers)
Product number: EL153B; GTIN: 00643169157248; Lot Number: 0009541707
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- InflateFX™ Gen II Fracture Management EL153BMedtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND1 recall0 reports
Other devices under the same FDA submission
36 devices cleared under K010246, K041454, K981251 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- KYPHON Xpander™ Inflatable Bone Tamp K09AMedtronic Sofamor Danek, Inc. · ARTHROSCOPE1 recall41 reports
- KYPHON Xpander KyphoPak Tray KPT1502Medtronic Sofamor Danek, Inc. · Arthroscope45 reports
- KYPHON Xpander Inflatable Bone Tamp K09AMedtronic Sofamor Danek, Inc. · Arthroscope35 reports
- KYPHON Xpander™ Inflatable Bone Tamp K09AMedtronic Sofamor Danek, Inc. · Arthroscope21 reports
- KYPHON Xpander™ Inflatable Bone Tamp K09AMedtronic Sofamor Danek, Inc. · ARTHROSCOPE17 reports
- KYPHON Xpander Inflatable Bone Tamp K08AMedtronic Sofamor Danek, Inc. · Arthroscope11 reports
- KYPHON Xpander™ Inflatable Bone Tamp K09AMedtronic Sofamor Danek, Inc. · ARTHROSCOPE9 reports
- KYPHON Xpander™ Inflatable Bone Tamp K08AMedtronic Sofamor Danek, Inc. · Arthroscope9 reports
- KYPHON Xpander™ Inflatable Bone Tamp K08AMedtronic Sofamor Danek, Inc. · ARTHROSCOPE8 reports
- KYPHON Xpander Inflatable Bone Tamp K13AMedtronic Sofamor Danek, Inc. · Arthroscope7 reports
- KYPHON Express Inflatable Bone Tamp K15BMedtronic Sofamor Danek, Inc. · Arthroscope7 reports
- KYPHON Xpander™ Inflatable Bone Tamp K08AMedtronic Sofamor Danek, Inc. · ARTHROSCOPE7 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 85611).
- Z-2390-2020 · TerminatedMedtronic Bone Void Filler, Small - Product Usage: is an injectable, moldable and biocompatible calcium phosphate bone void filler. The single-use Medtronic Bone Void Filler Kit c…
- Z-2391-2020 · TerminatedMedtronic Bone Void Filler, Large - Product Usage: is an injectable, moldable and biocompatible calcium phosphate bone void filler. The single-use Medtronic Bone Void Filler Kit c…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page