Skip to content

PerkinElmer 226 Sample Collection Device recallZ-2264-2020 · PerkinElmer Health Sciences, Inc. · initiated Dec 19, 2017

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from PerkinElmer Health Sciences, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

PerkinElmer 226 Sample Collection Devices Lot Number 40020001 had serial numbers printed on the pack labels that did not match the serial number range of the devices contained within the pack.

FDA root cause category: Packaging process control

What the company is doing

PerkinElmer sent an Urgent Medical Device Correction letter dated 12/22/2017 by email to the sole consignee. PerkinElmer requested the consignee quarantine the product and contact Kerry Chunko, PerkinElmer Sr. Quality Manager for further information 864-299-8787. x 104 ([email protected]). The firm also requested the consignee inform all those affected by this issue in their organization and to return the recall response form and return it by fax to 864-299-8787 or as a scanned by email to [email protected].

Recall details

Recall number
Z-2264-2020
Status
Terminated
Initiated by the company
Dec 19, 2017
Recalling firm
PerkinElmer Health Sciences, Inc.
Quantity
164,952 devices
Product code
PJC
FDA submissions
K121864
Distribution
US distribution in the state of Ohio

Product description

PerkinElmer 226 Sample Collection Device

Code information (models, lots, serial numbers)

UDI: (01)50812589020011(17)200831(10)40020001; Serial Numbers: 06879601-06879700

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page