A.L.P.S. PROXIMAL HUMERUS PLATING SYSTEMHIGH PLATE RIGHT recallZ-2323-2020 · Biomet · initiated May 14, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Distributed to the field without having completed design history files or design transfer activities
FDA root cause category: Device Design
What the company is doing
Zimmmer Biomet issued Urgent Medical Device Recall letter , dated May 14, 2020 to Risk Manger, Distributors,Surgeons, stating reason for recall, health risk and action to take: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form will be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule.
Recall details
- Recall number
- Z-2323-2020
- Status
- Terminated
- Initiated by the company
- May 14, 2020
- Recalling firm
- Biomet
- Quantity
- 1 OUS
- Product code
- HRS
- FDA submissions
- K143697
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of NY, GA, FL, MO and the countries of Canada, AUSTRALIA, CHILE, CHINA, JAPAN, MALAYSIA, NETHERLANDS.
Product description
A.L.P.S. PROXIMAL HUMERUS PLATING SYSTEM HIGH PLATE RIGHT 11 HOLE 197MM, Item Number: 110030408 - Product Usage: Intended for temporary internal fixation and stabilization of osteotomies and fractures during the normal healing process.
Code information (models, lots, serial numbers)
Lot Number: 408920 (01)00887868134883(17)270627(10)408920
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- A.L.P.S.® PROXIMAL HUMERUS PLATING SYSTEMHIGH PLATE RIGHT 110030408Biomet Orthopedics, LLC · Plate, fixation, bone1 recall0 reports
Product lines
Other devices under the same FDA submission
25 devices cleared under K143697 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- A.L.P.S.® PROXIMAL HUMERUS PLATING SYSTEM3.2MM LOCKING PEG 110025460Biomet Orthopedics, LLC · Plate, fixation, bone1 recall0 reports
- A.L.P.S.® PROXIMAL HUMERUS PLATING SYSTEMLOW PLATE LEFT 110030109Biomet Orthopedics, LLC · Plate, fixation, bone1 recall0 reports
- A.L.P.S.® PROXIMAL HUMERUS PLATING SYSTEMLOW PLATE RIGHT 110030208Biomet Orthopedics, LLC · Plate, fixation, bone1 recall0 reports
- A.L.P.S.® PROXIMAL HUMERUS PLATING SYSTEMLOW PLATE RIGHT 110030209Biomet Orthopedics, LLC · Plate, fixation, bone1 recall0 reports
- A.L.P.S.® PROXIMAL HUMERUS PLATING SYSTEM3.2MM CALIBRATED DRILL BIT 110017737Biomet Orthopedics, LLC · Plate, fixation, bone1 recall0 reports
- A.L.P.S.® PROXIMAL HUMERUS PLATING SYSTEMHIGH PLATE RIGHT 110030407Biomet Orthopedics, LLC · Plate, fixation, bone1 recall0 reports
- A.L.P.S.® PROXIMAL HUMERUS PLATING SYSTEMHIGH PLATE RIGHT 110030409Biomet Orthopedics, LLC · Plate, fixation, bone1 recall0 reports
- A.L.P.S.® PROXIMAL HUMERUS PLATING SYSTEMHIGH PLATE LEFT 110030307Biomet Orthopedics, LLC · Plate, fixation, bone1 recall0 reports
- A.L.P.S.® PROXIMAL HUMERUS PLATING SYSTEMHIGH PLATE LEFT 110030308Biomet Orthopedics, LLC · Plate, fixation, bone1 recall0 reports
- A.L.P.S.® PROXIMAL HUMERUS PLATING SYSTEMDRILL BLOCK SET SCREW 110017551Biomet Orthopedics, LLC · Plate, fixation, bone1 recall0 reports
- A.L.P.S.® PROXIMAL HUMERUS PLATING SYSTEMDRILL BLOCK GUIDE SLEEVE 110017553Biomet Orthopedics, LLC · Plate, fixation, bone1 recall0 reports
- A.L.P.S.® PROXIMAL HUMERUS PLATING SYSTEMDRILL BLOCK GUIDE SLEEVE 110017563Biomet Orthopedics, LLC · Plate, fixation, bone1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 85701).
- Z-2305-2020 · TerminatedA.L.P.S. PROXIMAL HUMERUS PLATING SYSTEM DRILL BLOCK SET SCREW, Item Number: 110017551 - Product Usage: Intended for temporary internal fixation and stabilization of osteotomies an…
- Z-2306-2020 · TerminatedA.L.P.S PROXIMAL HUMERUS PLATING SYSTEM DRILL BLOCK GUIDE SLEEVE 3.2MM, Item Number: 110017553 - Product Usage: Intended for temporary internal fixation and stabilization of osteot…
- Z-2307-2020 · TerminatedA.L.P.S. PROXIMAL HUMERUS PLATING SYSTEM DRILL BLOCK GUIDE SLEEVE 2.7MM, Item Number: 110017563 - Product Usage: Intended for temporary internal fixation and stabilization of osteo…
- Z-2308-2020 · TerminatedA.L.P.S. T15 LOW PROFILE CORTICAL SCREW 3.5mm x 28mm, Item Number: 110017628 - Product Usage: Intended for temporary internal fixation and stabilization of osteotomies and fracture…
- Z-2309-2020 · TerminatedA.L.P.S. PROXIMAL HUMERUS PLATING SYSTEM 3.2MM CALIBRATED DRILL BIT, Item Number: 110017737 - Product Usage: Intended for temporary internal fixation and stabilization of osteotomi…
- Z-2310-2020 · TerminatedA.L.P.S. T15 Multi-Directional Locking Screw 3.5mm x 60mm, Item Number: 110017960 - Product Usage: Intended for temporary internal fixation and stabilization of osteotomies and fra…
- Z-2311-2020 · TerminatedA.L.P.S. PROXIMAL HUMERUS PLATING SYSTEM 3.2MM LOCKING PEG 60MM , Item Number: 110025460 - Product Usage: Intended for temporary internal fixation and stabilization of osteotomies …
- Z-2312-2020 · TerminatedA.L.P.S. PROXIMAL HUMERUS PLATING SYSTEM PROXIMAL HUMERUS PLATE TEMPLATE 90MM, Item Number: 110025471 - Product Usage: Intended for temporary internal fixation and stabilization…
- Z-2313-2020 · TerminatedA.L.P.S. PROXIMAL HUMERUS PLATING SYSTEM PROXIMAL HUMERUS PLATE TEMPLATE 140MM Item Number: 110025472 - Product Usage: Intended for temporary internal fixation and stabilization …
- Z-2314-2020 · TerminatedA.L.P.S. PROXIMAL HUMERUS PLATING SYSTEM PROXIMAL HUMERUS PLATE TEMPLATE 234MM Item Number: 110025473 - Product Usage: Intended for temporary internal fixation and stabilization …
- Z-2315-2020 · TerminatedA.L.P.S. PROXIMAL HUMERUS PLATING SYSTEM LOW PLATE LEFT 11 HOLE 190MM, Item Number: 110030108 - Product Usage: Intended for temporary internal fixation and stabilization of osteoto…
- Z-2316-2020 · TerminatedA.L.P.S. PROXIMAL HUMERUS PLATING SYSTEM LOW PLATE LEFT 14 HOLE 227MM, Item Number: 110030109 - Product Usage: Intended for temporary internal fixation and stabilization of osteoto…
- Z-2317-2020 · TerminatedA.L.P.S. PROXIMAL HUMERUS PLATING SYSTEM LOW PLATE RIGHT 11 HOLE 190MM, Item Number: 110030208 - Product Usage: Intended for temporary internal fixation and stabilization of osteo…
- Z-2318-2020 · TerminatedA.L.P.S. PROXIMAL HUMERUS PLATING SYSTEM LOW PLATE RIGHT 14 HOLE 227MM, Item Number: 110030209 - Product Usage: Intended for temporary internal fixation and stabilization of osteot…
- Z-2319-2020 · TerminatedA.L.P.S. PROXIMAL HUMERUS PLATING SYSTEM HIGH PLATE LEFT 7 HOLE 140MM, Item Number: 110030307 - Product Usage: Intended for temporary internal fixation and stabilization of osteoto…
- Z-2320-2020 · TerminatedA.L.P.S. PROXIMAL HUMERUS PLATING SYSTEM HIGH PLATE LEFT 11 HOLE 197MM, Item Number: 110030308 - Product Usage: Intended for temporary internal fixation and stabilization of osteot…
- Z-2321-2020 · TerminatedA.L.P.S. PROXIMAL HUMERUS PLATING SYSTEM HIGH PLATE LEFT 14 HOLE 234MM, Item Number: 110030309 - Product Usage: Intended for temporary internal fixation and stabilization of osteot…
- Z-2322-2020 · TerminatedA.L.P.S. PROXIMAL HUMERUS PLATING SYSTEM HIGH PLATE RIGHT 7 HOLE 140MM, Item Number: 110030407 - Product Usage: Intended for temporary internal fixation and stabilization of osteot…
- Z-2324-2020 · TerminatedA.L.P.S. PROXIMAL HUMERUS PLATING SYSTEM HIGH PLATE RIGHT 14 HOLE 234MM, Item Number: 110030409 - Product Usage: Intended for temporary internal fixation and stabilization of osteo…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page